Parker SciLog LabTec Installation, Operating, & Maintenance Instructions page 23

Table of Contents

Advertisement

 
3. Press "C" again to select "DISP WEIGHT". Use the "A" key to increase the selected weight to
200.00 mL. Then press "C" to select.
4. Next, use the "A" and "B" keys again to scroll to "SNIFFLE" and press "C" to select. Use the "A"
key to increase this to 0.3, and press "C" to select.
5. All other parameters will be left at their defaults. Press the "EXIT" key to return to the "Weight"
screen and Press the "Exec" key, this will show theExec program that matches the Edit
program you were just in. Continuing with this example, press "C" to select, and the LabTec™
will initialize the scale and show the following display:
6. Press the "RUN" key, or the foot switch if you are using one, and the LabTec™ will Tare the
scale, dispense 200.00 gm, and show the following screen:
Where DW = Dispensed Weight, CW = Cumulative Weight, and SC = Sample # for the weight
dispensed, which will increment by one for each occurrence. The display will then alternate
between this screen and the one before it. Continue pressing the "RUN" key until you have
finished dispensing the needed number of aliquots.
 
WEIGHT
Incr.
Decr.
A
B
SNIFFLE
Incr.
Decr.
A
B
WEIGHT SET:
Press RUN when Ready
DW= 199.95 gm
199.9 gm
200.00gm
Select
C
0.3
Select
C
200.00 G
23