11.1.5 Group 5 - Error codes with five beeps every two seconds
2 s
Code
Name
11
ERROR_FLASH_ERROR
12
ERROR_INPUT_VOLTAGE_MEAS_ERROR
13
ERROR_INPUT_VOLTAGE_LOW
14
ERROR_INPUT_VOLTAGE_HIGH
15
ERROR_INPUT_FREQUENCY_LOW
16
ERROR_INPUT_FREQUENCY_HIGH
17
ERROR_INPUT_PHASE_MISSING
18
ERROR_UNSUPPORTED_PCB_VER
19
ERROR_STICKY_KEY

11.2 Other signal tones

User's manual
Hardware problem. See table 5.
Table 5: Error codes with five beeps every two seconds
Description
Auto position read or write failed. Try power
off/on or replace PCB.
Input AC voltage not detected. PCB is
powered from DC voltage or it is broken. Fix
mains input or replace PCB.
Input AC voltage is below minimum level.
Input AC voltage is above maximum level.
Input AC frequency is below minimum level.
Input AC frequency is above maximum level.
One of the 24 VAC phases is missing. Check
connections between transformer and PCB.
PCB version is either too old or new for current
software version. Update software or replace
PCB.
Some user interface button is stuck. Check
foot controls, additional keyboards and internal
wirings.
In the following figures a signal tone is shown with a note,
and the pitch of sound is expressed in Hz.
Hz
1479
1100
832.5
120ms
739.5
120ms
660
120ms
100
200
300
1.
The chair is switched on.
2.
An automatic chair position has been programmed
into the memory.
11 CHAIR SIGNAL TONES
t/ms
400
500
600
Planmeca Chair
25