Auto Sequence Testing Function Description; Fig.3-3 Auto Seq Function Operation Flow Chart - Prodigit 3260A Series Operation Manual

High power load
Table of Contents

Advertisement

PRODIGIT

3-3 AUTO SEQUENCE testing function description

There are two modes in AUTO SEQUENCE function, EDIT MODE and TEST MODE, The
AUTO SEQ mode can be entered by press S3 + S4 key simultaneously, then press STORE
key again to enter the EDIT MODE, or press START key to enter the TEST MODE, Please
refer to the flow chart operation below:
3.3.1
EDIT MODE
The TEST Mode of Auto-Sequence function is entered by press S3 + S4 key
simultaneously, The EDIT mode of Auto Sequence function is entered by press S3 +
S4 key simultaneously, the S3 and S4 LED are ON to indicate Auto-sequence mode,
then the Edit mode of Auto-sequence function is proceed by press STORE key.
The EDIT MODE flow chart is described below:
3.3.1.1
3.3.1.2
3.3.1.3
3.3.1.4
Press S3+S4 key
Press STORE key
STORE MODE
Press Exit key

Fig.3-3 AUTO SEQ function operation flow chart

There are nine Auto Sequence (n1-n9) can be edit within 3260A Series
Each Auto Sequence has up to 16 Test step, where each step is any one
memory of 5 sets Store memory
Each test step has t1 (test time) and t2 (delay time), the unit is 100mS, the
range is 0.1S - 9.9S in 100mS resolution. 3260A series load mainframe will
check each module GO/NG at the end of t1 (test time), the next step will be
started after duration t2 (delay time).
The test step sequence can be up to 16 step, and can be terminated by
press EXIT key if less than 16 step is required.
Display 3260 Serie high ower electronic
Show r1-r9
Firmware version r1-r9
Normal Mode
bn
Display bank
Display bank no.
Show n1~n9
AUTO SEQ.
Press STATE key
STATE MODE
3260A Series Operation Manual 21