C-7. CL-207-E Modbus Address Mappings (Base 1)

Address Description
30001 ~
Analog input value of channel 0 to 5. channel 0: HCHO in 1ppb,
30006
channel 1: relative humidity in 0.01%, channel 2: temperature in
40001 ~
0.01°C, channel 3:temperature in 0.01°F, channel 4: dew point
40006
temperature in 0.01°C, channel 5: dew point temperature in 0.01°F
40225 ~
High alarm limit of channel 0 to 5, channel 0: HCHO in 1ppb,
40230
channel 1: relative humidity in 0.01%, channel 2: temperature in
0.01°C, channel 3:temperature in 0.01°F, channel 4: dew point
temperature in 0.01°C, channel 5: dew point temperature in 0.01°F
40235 ~
Low alarm limit of channel 1 to 5, channel 1: relative humidity in
40238
0.01%, channel 2: temperature in 0.01°C, channel 3:temperature
in 0.01°F, channel 4: dew point temperature in 0.01°C, channel 5:
dew point temperature in 0.01°F
40272
Modbus NetID
Only for Modbus TCP protocol
30301
Number of the digital input channels
40301
Only for Modbus TCP protocol
30311
Number of the digital output channels
40311
Only for Modbus TCP protocol
30321
Number of the analog input channels
40321
Only for Modbus TCP protocol
30331
Number of the analog output channels
40331
Only for Modbus TCP protocol
30352
Firmware version in hex format
40352
Only for Modbus TCP protocol
40449
HCHO offset in 1 ppb
40450
Relative humidity offset in 0.01%
40451
Temperature offset in 0.01°C
40481
Firmware version (low word)
40482
Firmware version (high word)
40483
Module name (low word), 0x0207
40484
Module name (high word), 0x434C
40485
RS-485 module address, 1 to 247
Only for Modbus RTU protocol
C-7. CL-207-E Modbus Address Mappings (Base 1)
Attribute
R
R/W
R/W
R/W
R
R
R
R
R
R/W
R/W
R/W
R
R
R
R
R/W
- 186 -