Prime The Tubing; Unlocking The Pump - Smiths Medical CADD-Solis Quick Reference Manual

Hide thumbs Also See for CADD-Solis:
Table of Contents

Advertisement

Prime the tubing

Ensure that the pump is stopped, the tubing is
disconnected from the patient, and the clamp
is open on the patient tubing.
1.
Press
Tasks
Prime Tubing option, then press S.
2.
When prompted, unlock the pump.
3.
Press
Prime
4.
Press
Stop Priming
the delivery will stop automatically at 10 ml.

Unlocking the pump

You can start and stop the pump, and deliver a PCA dose
(if one is programmed) without unlocking the pump.
For all other tasks, you must enter the security code.
When the security code is required,the pump displays
Enter Code to Continue.
1.
Use U and D until the first digit of the
code is shown on the screen.
2.
Press S to advance to the next digit.
3.
Repeat with the second and third digits,
then press S or
entering the code.
Note: If the pump displays Use Key or Code to Continue, you can
unlock the keypad using the pump key, or enter the code, as above.
use U and D to highlight the
to start priming.
when the air is removed, or
Accept Value
to finish