Change the program with the pump stopped

You can change settings when the pump is stopped. Before
you restart the pump, the screen displays all the patient
specific parameters which you must review and accept.
1.
Use U and D to highlight the patient specific
parameter you want to change, for example,
Continuous Rate and press S.
2.
Unlock the keypad. The screen shows the
current setting of your selected parameter.
For rates, the range of soft limits for the
protocol are shown in green, and the range of
hard limits are in amber.
3.
Use U and D to set the new value and
press
Confirm.
4.
If a soft limit is exceeded, you see the message
Confirm soft limit override? Press
5.
When the change is complete, press T
Review pump settings. Press
6.
Carefully check the patient specific parameters. Press Accept
7.
Value on each parameter, or press S to change.
8.
When you have finished the review, press
Start Pump? Press
9.
The Home screen is displayed, showing the
Reports
and
Yes
to start the infusion.
Tasks
menu keys.
Yes.
Review.
Next.