Changing medication cassette / administration set
Follow your standard protocol for preparing the cassette.
Press T. Pump displays Stop pump?
1.
2.
Press
Yes.
3.
Clamp the tubing and disconnect from the patient.
4.
Unlock the cassette latch and remove the empty cassette.
5.
Attach a new cassette (see Attaching cassette /
administration set).
Set Reservoir Volume? press
6.
7.
Enter security code (see Unlocking the pump).
8.
Press S to reset to the protocol volume
displayed, or U and D to adjust.
9.
When the reservoir volume is correct,
press
Save.
volume.....saving.
10.
Prime the tubing when prompted (see Prime the tubing).
Review pump setting and start pump? Press
11.
12.
Press Review, then check the patient specific parameters and
Accept Value on each parameter.
13.
When you have finished the review, press
pump? Press
Pump displays Pump is stopping.
Pump displays Reservoir
Yes.
Yes.
Next
Yes.
to continue.Start