Sony MCC-1000MD Instructions For Use Manual page 8

Hide thumbs Also See for MCC-1000MD:
Table of Contents

Advertisement

Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.7 GHz
NOTE 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the MCC-1000MD is used
exceeds the applicable RF compliance level above, the MCC-1000MD should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the MCC-1000MD.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
c The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are
6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and
40.66 MHz to 40.70 MHz.
3 V/m
IEC 60601-1-2: 2007
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
IEC 60601-1-2: 2014
d = 2.0 √P 80 MHz to 2.7 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
should be less than the
compliance level in each
frequency range.
Interference may occur in the
vicinity of equipment marked
with following symbol:
8
a
b