Sony MCC-1000MD Instructions For Use Manual page 5

Hide thumbs Also See for MCC-1000MD:
Table of Contents

Advertisement

Important EMC notices for use in medical environments
• The MCC-1000MD needs special precautions regarding EMC and needs to be installed and
put into service according to the EMC information provided in the instructions for use.
• The MCC-1000MD is intended for use in a professional healthcare facility environment.
• Portable and mobile RF communications equipment, such as cellular phones, can affect the
MCC-1000MD.
Warning
• Portable RF communications equipment should be used no closer than 30 cm (12 inches) to
any part of the MCC-1000MD. Otherwise, degradation of the performance of this
equipment could result.
• If the MCC-1000MD will be used adjacent to or stacked with other equipment, normal
operation of the MCC-1000MD under such configurations should be verified via
observation.
• The use of accessories and cables other than those specified, with the exception of
replacement parts sold by Sony Corporation, may result in increased emissions or decreased
immunity of the MCC-1000MD.
Type of cable
CCMC-SA15 cable
CCMC-EA05 cable
Guidance and manufacturer's declaration – electromagnetic emissions
The MCC-1000MD is intended for use in the electromagnetic environment specified below. The
customer or the user of the MCC-1000MD should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
List of cables used for EMC test
Compliance
The MCC-1000MD uses RF energy only for
its internal function. Therefore, its RF
Group 1
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
The MCC-1000MD is suitable for use in all
Class B
establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
Class A
domestic purposes.
Complies
5
Specifications
15 m, shielded
5 m, shielded
Electromagnetic environment – guidance

Advertisement