Guidance and manufacturer's declaration – electromagnetic immunity
The MCC-1000MD is intended for use in the electromagnetic environment specified below. The
customer or the user of the MCC-1000MD should assure that it is used in such an environment.
Immunity test
Electrostatic
±8 kV contact
discharge (ESD)
IEC 61000-4-2
±15 kV air
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/
output lines
Surge
±1 kV line(s) to
line(s)
IEC 61000-4-5
±2 kV line(s) to
earth
Voltage dips, short
0% U
interruptions and
(100% dip in U
voltage variations
for 0.5/1 cycles
on power supply
input lines
40% U
(60% dip in U
IEC 61000-4-11
for 5 cycles
70% U
(30% dip in U
for 25/30 cycles
(for 0.5 sec)
0% U
(100% dip in U
for 250/300 cycles
(for 5 sec)
Power frequency
30 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
a For example, 10/12 means 10 cycles at 50 Hz or 12 cycles at 60 Hz.
IEC 60601
Compliance
test level
±8 kV contact
±15 kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV
differential mode
±2 kV common
mode
0% U
T
)
(100% dip in U
T
a
for 0.5/1 cycles
40% U
T
)
(60% dip in U
T
for 5 cycles
70% U
T
)
(30% dip in U
T
a
for 25/30 cycles
(for 0.5 sec)
0% U
T
)
(100% dip in U
T
a
for 250/300
cycles
(for 5 sec)
30 A/m
Electromagnetic environment –
level
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
a relative humidity of at least
30% is recommended.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
T
)
that of a typical commercial or
T
a
hospital environment. If the user
of the MCC-1000MD requires
continued operation during
T
)
power mains interruptions, it is
T
recommended that the
MCC-1000MD be powered from
an uninterruptible power supply
T
)
or a battery.
T
a
T
)
T
a
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.