To adjust the calibration of the AC Current Probe (refer to Figure 4 and 5):
1. Connect the AC Current output of the calibrator to the 50-turn coil.
2. Turn the instruments on and allow a 30-minute warm up period.
3. Connect the Current Probe to the digital multimeter (DMM) input via the BNC
adapter.
4. Select the 1 V AC range on the DMM.
5. Select the 1000 A range on the Current Probe.
6. Set the Calibrator to output 10 A @ 50 Hz into the 50 turn coil, thus producing 500 A
turns.
7. With the DMM in the 1 V AC range and the Current Probe centered on the coil,
adjust the potentiometer to read 0.500V +/- 1 mV.
8. Put the calibrator into standby mode.
5500A/coil 50 turns
Black
Black
HEWLETT
3458A
PACKARD
MULTIMETER
FUNCTION / RANGE
ACDCV
OHMF
ACDCI
Power
DCV
ACV
OHM
DCI
On
Off
Range
Auto
Hold
Test
Reset
Address
Menu Scroll
Display / Window
Figure 5. Calibration and Verification Setup
AC Current Probe for Oscilloscopes
5520A calibrator
Red
HP 3458A
MENU
NUMERIC / USER
C
NULL
STAT
E
Auto
NPLC
7
8
Cal
L
N
MEAN
SDEV
Auto
Offset
4
5
Zero
Comp
R
S
LF OFF
SCALE
PER
Trig
N Rdga/
1
2
ACI
FREQ
Trig
T
?
SYNC ACV
.
Recall
Store
0
Local
State
State
BNC/banana
adapter
Calibration Adjustment
Terminals
Sense
Input
Front
MATH OFF
Menu
(4 Wire)
(2 Wire)
Rear
9
E
HI
HI
UPPER
Error
200Vpk
1000Vpk
6
-
MAX
MAX
LO
LO
LOWER
Clear
Guard
Back
Open
3
200Vpk
To LO
Space
Guard
MAX
Recall
I
Enter
'
500Vpk
1A/250V
MAX
All Term 1000pk Max
akp02f.eps
11

Advertisement

Table of Contents
loading

Table of Contents