1. Installation Instructions

Tools Required

1. Allen wrench set, 1/8" 5/16"
2. Socket set with extension, 1/2", 9/16", 7/1", 3/4", 5/8"
3. Open end or closed end wrench, 1/2", 9/16", 7/1", 3/4", 5/8"
4. Flat head and Phillips head screwdriver
5. Tin snips
6. Razor blade

Installation Procedure

1. Uncrate contents from the boxes.
2. Inspect for any signs of damage.
3. Place the chair T-base (larger T-base) in the area of choice. Be aware that the chair
must recline and that, for the Pro model, the chair is 220VAC actuated.
4. Place the dynamometer T-base (smaller T-base) approximately 6" in front of the
chair T-base (do not align).
5. With a 1/8" Allen wrench, remove the screws from the chair T-base cover plate.
6. Remove the smaller bolts (1/2" socket).
7. Remove the larger bolts (3/4" socket).
8. Slide the dynamometer T-base into place and align the holes.
9. Install the T-Base bolts that go in toward the floor with a 1/2" socket.
NOTE: It may be necessary to keep the bolts loose until the bolts going into the side of the
dynamometer T-Base are aligned. Make certain to go back and tighten the bolts after the
front T-Base bolts are installed.
10. Tighten the bolts going into the side of the T-base with the 3/4" socket.
11. Remove the chair from the pallet with a 5/16" Allen wrench.
12. Place the chair on top of the chair T-base. Align the holes and use a 5/16" Allen
wrench to tighten. Be careful not to slide the chair around too much as this may
cause the paint to scratch.
13. Remove the dynamometer and gimbal from the pallet with a 5/16" Allen wrench.
14. Place the dynamometer and gimbal on the T-base and tighten with a 5/16" Allen
wrench. Ensure that the power cords are facing forward (away from the chair).
15. Unpack the computer, monitor, printer, keyboard, mouse, mouse pad, and hand-
held Comfort Stop.
16. Plug the pause switch into the connector located on the CDS midway up on the
right side.
17. Connect the Comfort Stop to the dynamometer.
18. Install the monitor (refer to the Monitor Install Procedure.) The power and signal
cable are already run and connect to the monitor.
19. Install the computer, located on the bottom shelf in the front of the CDS. Remove
the side covers (lift up and pull out and down) and slide the back of the computer
in from the left side. The cables are already run and ready to plug into the
computer as labeled.
20. Place the printer on the middle shelf. The cables are already run and located on the
right side of the shelf.
21. Plug in the power cord and connect the printer cable into the back of the printer.
22. Install the Motor Power cord and the Dyna Data cord into their respective, labeled,
connectors at the rear of the CDS cart.
23. Install cabling at the back of the T-Base (refer to AC Power Installation procedure.)