Page 1
BIODEX MULTI-JOINT SYSTEM INSTALLATION INSTRUCTIONS 840-000 840-000-10 840-000-30 850-000 850-000-10 850-000-30 FN: 08-051-CLR Rev A 9/19...
Page 2
Additional information and resources, including installation instructions, are available upon request or directly from the Biodex website, http://www.biodex.com. If the desired information is not found, please feel free to contact a local distributor or Biodex directly at supportservices@biodex.com. Thank you, Biodex Medical Systems, Inc.
Definition of Symbols The following symbols and their associated definitions are used and implied throughout this manual.
Page 5
Site Requirements Electrical Specifications System Power Requirements: Provide 220 VAC/60 Hz 20 amp dedicated electrical service with dedicated ground. Locate within ten feet of ear of the T-base. Receptacle will be a single Hubbell #IG-5461, rated for 250 VAC, or hospital grade equivalent.
1. Installation Instructions Tools Required 1. Allen wrench set, 1/8" 5/16" 2. Socket set with extension, 1/2", 9/16", 7/1", 3/4", 5/8" 3. Open end or closed end wrench, 1/2", 9/16", 7/1", 3/4", 5/8" 4. Flat head and Phillips head screwdriver 5.
Page 7
Figure 1.1. Multi-Joint System Cart and Monitor from the Rear...
Monitor Install Procedure Figure 1.2. Rear View of the CDS Cart Cabling for CDS Power Receptacle CDS Cart Power Seat Lift Power 195 to 245 Volt AC Receptacle Receptacle Input Receptacle Figure 1.3. Power Receptacles on the Back of Seat.
Page 9
Figure 1.4. Remove Screws from the CDS Cart Power Receptacle. Figure 1.5. Loosen the Clamp on the Cable. Strain Relief Slot Facing Up Cable Clamp Strain Relief Screw Facing Left Figure 1.6. Cable Positioned at CDS Cart Power Receptacle.
Seat Lift Power Cord Cable Clamp Screw Cable Strain Relief Figure 1.10. Installing the Seat Lift Power Cord. AC Input Receptacle Power Connection Figure 1.11. Loosen the Clamp on the Cable. Figure 1.12. Loosen the Cable Clamp Screw and Remove the Strain Relief.
Page 12
Cable Clamp Screw Figure 1.13. Removing the Screws from the AC Input Receptacle. Figure 1.14. Installing the AC Input Cable.
Date of Installation: Site Information: (Print clearly or type) FACILITY: ADDRESS: CITY, STATE: CONTACT AT SITE: PHONE: ( FAX: ( Biodex Representative: Installation Start Time: Serial Number: Model Number: Software Version: Firmware Version: Biodex Contacts: 1-800-224-6339 Software Support-ext. 2120 Technical Support-ext. 2116 Technical and Software Support-ext.
Page 14
System Site Checkout 1. Unpack and assemble system. Check for physical damage or cosmetic defects. Comments: 2. In "setup" use knee accessory and set the range limits maximum (total ROM- degrees is between 149-152). Comments: 3. While in "Passive" mode, test percent ROM controls, cushion and the pause controls.
Page 15
Description Customer request/comments: With the exception of any items listed above, the Biodex System has been completely installed satisfactorily and is functioning properly. The Biodex representative has instructed the facility on startup, hardware and software. The facility is capable of proper operation of the Biodex System.