SAnItAtIon
9.1
Sanitising products
ATTENTION! The sanitising agents are corrosive.
The best sanitising and disinfecting agents are those used more commonly in the
industrial sector. Always comply with the manufacturers instructions supplied for the
specific application. If possible, ask the manufacturer to guarantee the corrosion level of
the solutions used.
Any changes to these features could damage the item.
It is highly important to follow the specifications on concentration, temperature and the
reaction times. Any changes to these features could damage the machine.
During the sanitation phases, only use:
▪ cold mineral water;
▪ hot water max. 95°C;
▪ alkaline solutions max. 80°C;
▪ acid solutions;
▪ disinfectant solutions.
Do not use sulphuric acid or mineral acids such as HCl, H
9.2
Sanitation with products containing halogens
ATTENTION! Do not use products containing halogens when sterilising the closed circuit,
the bed could be damaged.
If used incorrectly, these products can corrode the steel, specifically if the pH is low.
Accurately check before using these solutions.
If the machine has to be sanitised using sanitising products containing halogens (e.g.
chlorine), follow these instructions:
▪ the pH must be over 10;
▪ the temperature must not exceed 40°C;
▪ the solution must not stay in contact with the bed for more than 20 min.;
▪ use a max concentration of 50 ppm of active chlorine;
▪ after sanitation, rinse with plenty of water.
9.3
Sanitation with acid products
ATTENTION! After each sanitation cycle with acid products, the bed must be immediately
and well rinsed with water.
9.3.1
Chlorinated products
ATTENTION! Ensure the chlorinated concentration in the chemical washing solution is
lower than 50 ppm. If the concentration exceeds the previous value, the parts in contact
with the product could become corroded.
9.3.2
Chemical products that release oxygen
ATTENTION! These agents could produce reactions with the gasket and rubber parts of
the bed and could destroy them during prolonged treatment with a closed circuit (not
recommended). You are advised to only carry out brief disinfection of this type before
operating the bed.
SO
, HNO
2
4
3