Night Safety Position - PamMobility SYNAGO 727T0045 Instructions For Installation And Use Manual

Adjustable height bed
Hide thumbs Also See for SYNAGO 727T0045:
Table of Contents

Advertisement

48

7.17 Night safety position

The bed is equipped with a "night safety" position, or rather a function that automatically
brings the surface of the bed to the minimum height of 380 mm from the floor.
This position in obtained using an electrical actuator controlled by a control with a
magnetic key, from the Supervisor control panel and the external side rail control panels,
exclusively for caregiver use.
ATTENTION! The night safety position can only be activated if the command locking
selector is in the "release all commands" position".
To bring the bed to the programmed "night safety" position, proceed as follows:
WITH THE FREE PUSH BUTTON PANEL
▪ activate the position with the magnetic key (see paragraph 5.1 Free push button
panel);
▪ press button
1h:
the bed lowers to the standard minimum height;
press button 1h again
until the night safety position is completely reached.
Adjustable height bed Synago - ID01188 - Rev. 0.0 -
15/11/2015
1h

Advertisement