Eiki EIP-WSS3100 Service Manual page 55

Hide thumbs Also See for EIP-WSS3100:
Table of Contents

Advertisement

3). Full On/Full Off Contrast
Procedure
Criteria
4). Uniformity
Procedure
Criteria
- Full white pattern & Full black pattern
- Use C100 to measure brightness values of full
white pattern P5 & full black pattern B5 ( see
image: full white)
- Follow Contrast formula to calculate contrast
values.
☼ Contrast Formula
P5/B5
Note: P5 = Lux of center in full white pattern
B5 = Lux of center in full black pattern
• 1600:1
- Full white pattern
- Use CL100 to measure brightness values of
P1~P9 (see image: full white).
- Follow the Uniformity formula to calculate
average values.
☼ Uniformity Formula
JBMA Uniformity = Avg. (P1, P3, P7, P9) /
• 65%
P5 *100%
Full black pattern
Confidential
4-7