4.
Maintenance
If irregularities occur during operation, use the displayed error codes to determine
how to proceed further.
4.1.
Troubleshooting
Problem
Alternating
display
TMP/ERR
AD/ERR
EE/ERR
PRG/ERR
RAM/ERR
LED/ERR
VIB/ERR
UNK/ERR
/ERR
SNS/ERR
Instrument
does not turn
ON
*) If the error occurs during the warranty period, please contact the MSA customer service. Otherwise,
the instrument must be put out of operation.
FASIT Srl
- Via Orsera 13/b - 34145 TRIESTE - ITALY - Tel +39 040 825604 -
Description
Temperature out of range
Sensor feedback error
EEPROM error
Memory error
RAM error
LED error
Vibration alarm error
Indefinable error
Battery warning (no alarm)
Battery warning
(with alarm – flash LEDs, horn)
Sensor error
Low battery
MAINTENANCE
Reaction
Return instrument to normal
temperature range and
recalibrate.
*)
Inform MSA
Remove battery and
recalibrate.
*)
Inform MSA
*)
Inform MSA
*)
Inform MSA
*)
Inform MSA
Check alarm LEDs
Check vibration alarm
*)
Inform MSA
Remove from service as soon
as possible and replace battery
Instrument is no longer sensing
gas.
Remove from service and
replace battery.
*)
Inform MSA
Replace battery