4
Operation

Saving programs

Prewarming or precooling the centrifuge
For setting the pretemp value for the centrifuge proceed as follows:
1. Press the
menu.
The display shows the message "Set PreTemp
"
2. Enter the desired value by pressing the
3. Close the centrifuge lid.
The display shows the current temperature inside the rotor chamber.
4. Press the "START" key
The rotor chamber is cooled down or heated up to the preset temperature.
5. Press the "STOP" key
The display shows the current temperature inside the rotor chamber.
Bucket selection
Bucket selection is only possible for swing-out rotors. The bucket code corresponds to the last four
digits of the bucket catalog number.
1. Press the "BUCKET" key.
The display shows the following message:
2. Press the BUCKET key repeatedly until the bucket being used is displayed.
Saving programs
1. Enter the program parameters.
2. Press any of the program store keys for 5 seconds.
4-6
SL 16 / 16R
key below the TEMPERATURE display in order to open the temperature selection
.
.
key repeatedly, until the desired value shows.