Regulatory Standards (Only For Sc42 With Amphenol Connector) - YOKOGAWA SC4A Instruction Manual

Specific conductivity sensors
Hide thumbs Also See for SC4A:
Table of Contents

Advertisement

SA11 Smart Adapter

3.1.6. Regulatory standards (only for SC42 with Amphenol connector)

- ATEX
Certificate no.
- IECEx
Applying standards
Certificate no.
Conformity
- Electrical data for ATEX/IECEx
Special conditions (X)
: Directly connected to the analyzer using a
WU11 cable up to 100 meters
or
Connected to a BA11 connetion box
using WU11 cable upto 100 m. The BA11
connection box is connected to the analyzer
using a WU11 cable up to 100m
: Directive 2014/34/EU
by applying:
EN 60079-0
EN 60079-11
EN 60079-26
: DEKRA 14ATEX0074 X
II 1 G Ex ia IIC T4... T6 Ga
: IEC 60079-0
: IEC 60079-11
: IEC 60079-26
: IECEx DEK 14.0032X
Ex ia IIC T4...T6 Ga
: EAC (Eurasia)
TS (Taiwan)
: For sensor input circuit connected to
A certified intrinsically safe circuit with the following
maximum values:
Ui = 14.4 V; Ii = 116.5 mA; Pi = 0.342 W
or
Certified intrinsically safe Yokogawa Contact
Conductivity transmitter Model FLXA21 series or
Model SC202S series.
The effective internal capacitance Ci and the
effective internal inductance Li of the sensor depends only
upon the properties and length of the integral cable.
: T6 for Tamb. -30°C to 40°C
T5 for Tamb. -30°C to 95°C
T4 for Tamb. -30°C to 130°C
15