3.4. Wiring SC41/SC42/SC49 sensors

The SC41, SC42 and SC49 sensors are provided with a fixed connector. The standard cable
used to connect the sensor with Amphenol connector to the analyser is the WU40. The standard
cable used to connect the sensor with VP connector to the analyzer is the WU10. These cables
are available up to 25/20 meters. When a longer cable run is necessary (maximum cable run
is 60 meters), this can be done by using the WF10 extension cable in combination with the
BA10 connection box. The connection of the WU40 cable, WU10 cable and WF10 cable to the
Yokogawa Contact Conductivity analyser are given in table 4.
Amphenol
16
15
connector
CELL
Variopin
connector
Fig. 27 Top view connector system SC41/
SC42 2-electrode

3.5 Modelcode SC41/SC42/SC49

Model
Suffix Code Option code
SC41 (SC49)
SC42
Materials
-E
-S
-F
-T
Mounting
P
V
Cell constant
Type
Options
Note: Suffix V not ATEX/IECEx certified. Suffix V not in combination with suffix -E (Epoxy).
11
12
temp.
13
14
element
1)
0
1
2
3
4
5
6
8
Amphenol
connector
CELL
Variopin
connector
Fig. 28 Top view connector system SC42/
SC49 4-electrode
Description
Conductivity Sensor 2- or 4- electrodes + Ni100
Conductivity Sensor 2- or 4- electrodes + Pt1000
Epoxy / graphite
Stainless steel AISI 316L / PEEK
PVDF / Glass / Platinum
PTFE / Glass / Platinum
Plug-in type, plug-socket connector
Plug-in type, VarioPin connector with SENCOM
ID-chip 1)
Cell constant
C = 1
cm-1
C = 0,1 cm-1
C = 0,01 cm-1
2-electrode, flow cell
2-electrode, insertion cell
2-electrode, insertion cell with DN25 collar
4-electrode, flow cell
N/A
25
11
16
12
15
temp.
13
14
element
IM 12D07J01-10E-E