6. Output Telegram Formats
If the optional pressure sensor is fitted and enabled then the only data format that supports this is the Valeport
NMEA format. A number of other formats have been implemented to provide compatibility with existing systems.
6.1. Valeport NMEA Telegram ($PRVAT)
$PRVAT, xx.xxx,M, xxxx.x, dBar*hh<CR><LF>
Field Number:
1. Range from transducer
2. Units
3. Pressure
4. Units
5. Checksum
Example:
$PRVAT,00.115,M,0010.1,dBar*39
$PRVAT,00.116,M,0010.0,dBar*3d
6.2. $SDDBT Telegram
$SDDBT,x.x,f,x.x,M,x.x,F*hh<CR><LF
Sounder, Depth: Depth below transducer
Field Number:
1. Depth, feet
2. f = feet
3. Depth, meters
4. M = meters
5. Depth, Fathoms
6. F = Fathoms
7. Checksum
Example:
$SDDBT,0.4,f,0.1,M,0.1,F*02
$SDDBT,0.4,f,0.1,M,0.1,F*02
Page 19
NB: This format limits the resolution of the altimeter to 0.1m
Output Telegram Formats