Guidelines to Avoid Measurement Errors

To avoid errors, see Table 25.
Sources of Inaccuracy
Thermal EMFs can cause series (Normal) mode
interference, particularly where large currents
have a heating effect at junctions. In otherwise
thermoelectrically-balanced measuring circuits,
cooling caused by drafts can upset the balance.
Table 25. Guidelines to Avoid Errors
Screen thermal junctions from drafts.
Reference Multimeter and 8 ½ Digit Multimeter
Guidelines to Avoid Measurement Errors
Avoiding or Minimizing Inaccuracy
Before taking readings allow time for thermal
equilibrium to be reached.
Use conductors, joints and terminals with a
good margin of current-carrying capacity.
Avoid thermoelectric junctions where
possible:
o Use untinned single-strand copper wire of
high purity.
o Avoid making connections through Nickel,
Tin, Brass and Aluminum. If oxidation is a
problem, use gold-plated copper
terminals, and replace the terminals
before the plating wears off.
o If joints must be soldered, low-thermal
solders are available, but crimped joints
are preferred.
o Use low-thermal switches and relays
where they form part of the measuring
circuit.
o Balance one thermal EMF against another
in opposition, where possible. (Switch and
relay contacts, terminals, etc.)
125