Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
Pole mount for PSM – Instructions for connecting to frame

3.3.5 Positioning PC boards

1
1
2
2
Figure 10
(1)
Mains power inlet
(2)
Battery 24 Vde input connector (F-CU8 - 12) / Display power outlet (F-CU8 rev. 03 to 09)
(3)
UPINET Board, B-UPI4NET
(4)
B-CPU6 and integrated display controller (primary/screen1)
(5)
Display Controller Board, B-DISPX (secondary/screen 2)
Display Controller Board, B-DISPX (3
(6)
Service reset button
(7)
Equipotential connector
26
.
Document no
M1137263-04
1. Attach the E-PSM(P) module to the Pole Mount.
2. Connect the cable of the Pole Mount for PSM to the F-CU8
with the E-INTPSM module (M1054424).
3. Check the module communication of the E-PSM(P)
module.
1. Attach the E-PSM(P) module to the Pole Mount.
2. Connect the cable of the Pole Mount for PSM to the S/5
F-CU5 with the Module Bus adapter for PSM (M1031025).
3. Check the module communication of the E-PSM(P)
module.
3
3
!
9
7
6
8
Rear view and positioning, F-CU8 (-12 shown)
4
4
5
6
7
5
rd
screen), Keyboard Interface Board, B-ARK