Temperature Calibrators Series TP 17...M Types TP 17165M • TP 17650M Please keep this operating manual for future reference. If the device is resold, please provide the operating manual along with it.
Dr. Siebert & Kühn GmbH & Co. KG Struthweg 7-9 • D - 34260 Kaufungen +49 5605 803-0 • +49 5605 803-555 info@sika.net • www.sika.net Hazard signs and other symbols used: CAUTION! Electric current! This sign indicates dangers which could arise from handling of electric current.
TP 17...M Device description Scope of delivery and accessories: Scope of delivery: Accessories (optional): Metal block calibrator according to order Transport case. data. Transport bag (only TP 17650M). Mains cable. Adapter sleeves. Sleeve exchange tools. ...
In order to guarantee that the device operates safely, the operator must act competently and be conscious of safety issues. SIKA provides support for the use of its products either personally or via relevant literature. The customer verifies that our product is fit for purpose based on our technical information.
Page 7
The calibrator cannot then be started any more. After it has cooled, send the calibrator for inspection to SIKA. The calibrator may not be used in an explosion-endangered atmosphere (ignitable or ex- plosive atmosphere).
TP 17...M Commissioning and operation Commissioning and operation CAUTION! Risk of injury or material damage! The calibrator can become very hot during operation. If the machine is operated without supervision, third-party persons in the vicinity could get injured. Moreover, flammable material could get into the machine and cause significant damage to property.
IMPORTANT! Mains connecting cable! The mains connecting cable may only be replaced by an equivalent cable. Use only original cables from SIKA or approved cables of the same type with the cor- rect design as replacements ( § 10.1 "Electrical characteristics").
Inserting: Only use the supplied adapter sleeves made of the appropriate material. If in doubt, contact SIKA for clarification. The appropriate adapter sleeve is inserted into the metal block with the aid of the sleeve exchange tool.
TP 17...M Commissioning and operation IMPORTANT! Mains failure or separation from the mains! If there is a mains failure, or if the main switch is turned off, or upon removal of the mains connection ("EMERGENCY STOP"), the built-in fan does not blow any more cooling air.
TP 17...M Operation 5.2 Calibration (calibration mode) As soon as the calibrator has been switched on, it switches to the calibration mode after ini- tialization. 5.2.1 Display of reference and set temperature The current reference temperature is displayed in the upper and the set temperature in the lower of the display.
Page 16
If the LED OUT 2 is not lit the fan is running at low speed. Fan / cooling control active b) Heating and cooling instrument (TP 17165M) The LED OUT 2 indicates the status of the output for the cooling control: •...
TP 17...M Operation 5.3 Set a set temperature (set point mode) In this operating mode it is possible to modify a saved set temperature. Press the P key shortly. The currently active set point memory, e.g. SP 1 (set point 1), is displayed by the upper display.
Page 19
TP 17...M Testing process / Calibration Performing testing process: During the testing process, one or more test points are approached and the reference tem- perature of the calibrator is compared with the measured temperature of the specimen. Set the set temperature to the value of the first test point. ...
Contact your supplier or directly SIKA. Please send the device for repair with a brief descrip- tion of the problem, the environmental conditions and the length of time the device was oper- ational before the problem occurred.
TP 17...M Troubleshooting 7.1 Return shipment to the manufacturer Please follow the instructions on the return of goods on our website (www.sika.net/en/services/return-of-products-rma.html). Technical changes reserved - 21 -...
The calibrator should, depending on the application situation, be inspected at appropriate in- tervals on the basis of DIN ISO 10012. We recommend you to return the calibrator to SIKA at intervals of max. 36 months or approx. 500 operating hours for recalibration and readjust- ment.
The calibrator of the series TP 17...M consists of various different materials. It must not be disposed of with household waste. Take the TP 17...M to your local recycling plant send the TP 17...M back to your supplier or to SIKA. * WEEE reg. no.: DE 25976360 - 24 -...
10 Technical Data The technical data of customised versions may differ from the data in these instructions. Please observe the information specified on the type plate. 10.1 Characteristics series TP 17…M Series TP 17165M TP 17650M Characteristics Kalibrator Temperature range -35…165 °C...
TP 17...M Technical Data 10.2 Heating and cooling periods The heating up and cooling times are dependent on the parameters of the test task used and the ambient conditions. The following times are guide values for a room temperature of 23 °C without settling processes.