Operating manual
Operating manual .................................................. page 1 - 60
Temperature calibrators
Series TP37 / TP3M
Types: TP 37200E.2 • TP 37165E.2 • TP 37166E.2
• TP 37700E • TP 3M165E.2 • TP 3M255E
Please keep this operating manual for future reference.
If the device is resold, please provide the operating manual along with it.
(Translation)

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TP 37166E.2 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for SIKA TP 37166E.2

  • Page 1 Operating manual ..........page 1 - 60 Temperature calibrators Series TP37 / TP3M Types: TP 37200E.2 • TP 37165E.2 • TP 37166E.2 • TP 37700E • TP 3M165E.2 • TP 3M255E Please keep this operating manual for future reference. If the device is resold, please provide the operating manual along with it.
  • Page 2: Table Of Contents

    Manage Function ....................31 5.4.3 Configure Function ....................32 Test specimen (DUT - Device under test) ..............33 5.5.1 Select DUT ......................33 5.5.2 Manage DUT ......................34 5.5.3 Configure DUT ....................... 34 - 2 - © SIKA • Ba_TP37-TP3M_en • 02/2018...
  • Page 3 10.1 Characteristics TP37 / TP3M ..................53 10.2 Heating up and cooling times ................... 55 10.2.1 TP 37200E.2 • TP 37165E.2 • TP 37166E.2 • TP 37700E (dry block) ....55 10.2.2 TP 37166E.2 (Infrared and surface) ..............55 10.2.3 TP 3M165E.2 ......................56 10.2.4 TP 3M255E......................
  • Page 4: About This Operating Manual

     Check the specified points or notices. mation that is marked with this symbol.  Reference to another section, document or source.  Follow the specified instructions and steps. Adhere to the given order. • Item. - 4 - © SIKA • Ba_TP37-TP3M_en • 02/2018...
  • Page 5: Device Description

    Further information on the series, the models of the individual types that can be supplied, and their fittings can be found in our catalogues at "catalogues.sika.net". Type plate: You find the type plate on the rear of the device. It includes the type designation, the serial number and the key electric specifications (...
  • Page 6: Delivery, Unpacking And Accessories

     Works certificate.  Magnetic stirrer with magnet lifter  Work cover with 5 Silicone plugs * only TP 37166E.2, TP 3M165E.2 and TP 3M255.E ** only TP 3M165E.2 and TP 3M255E IMPORTANT!  Use the type plate to check if the delivered unit corresponds to your order.
  • Page 7: Intended Use

    The calibrator is under guarantee for 12 months as from the date of delivery for construction errors or material defects. The guarantee is limited to repair or replacing the calibrator. SIKA also provides an extra 5-year guarantee for calibrators which are calibrated and in- spected annually by the SIKA DAkkS laboratory.
  • Page 8: Exclusion Of Liability

    In order to guarantee that the device operates safely, the operator must act competently and be conscious of safety issues. SIKA provides support for the use of its products either personally or via relevant literature. The customer verifies that our product is fit for purpose based on our technical information.
  • Page 9 The calibrator cannot then be started any more.  After it has cooled, send the calibrator for inspection to SIKA. Risk of injury from hazardous gases! When liquids are heated, the evaporation can result in dangerous gases being formed.
  • Page 10: Safety Instructions For The Application Of Calibration Liquids

    Mineral oil: SIKA supplies the calibrators only with silicone oil. Using mineral oil is possible, but must be done on your own responsibility. The danger and the risk must be borne by the customer and our warranty will be rendered null and void.
  • Page 11: Construction, Function And Test Task

    TP37 / TP3M Construction, Function and Test Task Construction, Function and Test Task 3.1 Construction  The calibrator consists of a robust, black-and-red painted steel housing and has a carry-  ing handle at the top The front part of the housing holds the electronics of the TP37 / TP3M for controlling the ref- ...
  • Page 12: Function

    The functions, test specimens and test tasks marked with the symbol are factory- defined. They can be neither deleted nor configured. They serve as templates for your own functions, test specimens and test tasks, which are marked with the symbol. - 12 - © SIKA • Ba_TP37-TP3M_en • 02/2018...
  • Page 13: Test Tasks

    TP37 / TP3M Construction, Function and Test Task 3.3 Test tasks Test tasks are containers for defined test conditions. They are helpful for recurring testing processes, for standardising test sequences and for generating measurement logs. All the required settings and configurations for the calibration of a test specimen are com- piled in a test task.
  • Page 14: Commissioning And Operation

    The plug of the mains connecting cable serves as a "KILL" switch.  Ensure that the plug is always easily accessible and easy to reach.  In an emergency, pull the plug, so that the calibrator is isolated from the mains. - 14 - © SIKA • Ba_TP37-TP3M_en • 02/2018...
  • Page 15: Electrical Connection

    IMPORTANT! Mains connecting cable! The mains connecting cable may only be replaced by an equivalent cable.  Use only original cables from SIKA or approved cables of the same type with the cor- rect design as replacements ( "Electrical characteristics").
  • Page 16: Preparing Of The Calibrator

    These characteristic values are stored as protected functions for the accompanying measur- ing inserts of the calibrator ( § 5.4). - 16 - © SIKA • Ba_TP37-TP3M_en • 02/2018...
  • Page 17: Dry Block Calibrator

    TP37 / TP3M Commissioning and operation 4.3.1 Dry block calibrator Type TP 37200E.2 TP 37165E.2 TP 37166E.2 TP 37700E TP 3M165E.2 TP 3M255E Dry block       Adapter sleeves with single or multiple holes are used for the calibration of straight temper- ature sensors.
  • Page 18: Calibrator For Infrared Thermometer

     Keep the measuring opening closed for as long as possible.  Only open the measuring opening briefly for measuring.  Existing ice or condensation can be removed by heating the infrared insert. - 18 - © SIKA • Ba_TP37-TP3M_en • 02/2018...
  • Page 19: Calibrator For Surface Temperature Sensors

    TP37 / TP3M Commissioning and operation 4.3.3 Calibrator for surface temperature sensors Type TP 37200E.2 TP 37165E2 TP 37166E.2 TP 37700E TP 3M165E.2 TP 3M255E Surface    A special surface insert is used for calibrating surface temperature sensors. This makes cal- ibration fast and easy.
  • Page 20: Micro Bath Calibrator

    ( § 5.2) or in the window "Configure Function" ( § 5.4.3 "Stirring speed"). IMPORTANT! Limited life part! The magnetic stirrer is a limited life part.  Replace worn-out magnetic stirrers. ( § 8.1 "Maintenance"). - 20 - © SIKA • Ba_TP37-TP3M_en • 02/2018...
  • Page 21 TP37 / TP3M Commissioning and operation Drain syringe and magnetic lifter: The drain syringe is used for pumping out the calibration liquid from the tank of the micro bath. The magnetic stirrer is removed with the help of the magnetic lifter. Both the activities are required before another adapter sleeve or calibration insert is inserted in the calibrator.
  • Page 22 The maximum fill level is ~0.45 litres. Tub insert: The maximum fill level with the tub insert is below the fixture for the sleeve exchange tool. The maximum fill level is ~0.32 litres. - 22 - © SIKA • Ba_TP37-TP3M_en • 02/2018...
  • Page 23 TP37 / TP3M Commissioning and operation 4.3.4.3 Filling the micro bath * IMPORTANT! Pay attention to § 4.3.4.2 "Information on filling amounts"!  When filling, leave enough room for thermal expansion, displacement by sensors and level rise due to stirring. ...
  • Page 24: Switching On, Cool Down And Switching Off

     Put on the main switch.  The fan of the device starts and the SIKA logo appears for ~ 40 s on the screen.  The type designation and the current software version are displayed for ~ 4 s.
  • Page 25 TP37 / TP3M Commissioning and operation Switching off: CAUTION! Before switching off, pay attention to the block temperature! The calibrator should first reach a safe block temperature before switching off. Upon switching off outside the safe temperature range, the calibrator can get damaged. ...
  • Page 26: Operation

    The operating concept of the TP37 / TP3M is based on pre-installed and self-defined test tasks for recurring testing processes. One-off testing processes can be directly set via the configuration range and then started. - 26 - © SIKA • Ba_TP37-TP3M_en • 02/2018...
  • Page 27: Main Window

    TP37 / TP3M Operation 5.1 Main window The individual icons and the most important ranges of the main window are presented in brief in the following.  Calibrator Set Up: By using the icon, you can reach the "Calibrator Set Up" window. There, you can change the settings of the calibrator (...
  • Page 28: Configuration Range

    It contains the icons available in this win- dow. Parameter range: The parameter range is the biggest part of the window. Information about the window and other operating elements are displayed in it. - 28 - © SIKA • Ba_TP37-TP3M_en • 02/2018...
  • Page 29: Operating Elements

    TP37 / TP3M Operation 5.3.1 Operating elements The following operating elements can be found in the parameter range of the window. Information box: Different kinds of information is displayed in the information box. This information can be explana- tions for the window, information on the selected element of the selection list, measurement values or measurement value curves.
  • Page 30: Toolbar And Icons

    For better differentiation and simpler operation, most of the parameter fields are marked with additional symbols before the text in the windows. These icons are also used in the op- erating manual. - 30 - © SIKA • Ba_TP37-TP3M_en • 02/2018...
  • Page 31: Functionality Of The Calibrator

    TP37 / TP3M Operation 5.4 Functionality of the calibrator Here, you can select and manage the function of the calibrator for the relevant measuring in- sert. You can create new functions or tailor unprotected functions to your requirements. IMPORTANT! When changing functions, please consider! The individual functions of the calibrator are independent.
  • Page 32: Configure Function

    Alignment values : If required, you can save the corresponding adjustment values for your own measuring in- serts here. Further information on this can be found under "Adjustment" ( § 8.4). - 32 - © SIKA • Ba_TP37-TP3M_en • 02/2018...
  • Page 33: Test Specimen (Dut - Device Under Test)

    TP37 / TP3M Operation Setting time  Input the new value in the input window and confirm the input.  The new value is adopted and you will be returned to the previous window. Stirring speed You can change the stirring speed here. The parameter field is only displayed with "Select Function"...
  • Page 34: Manage Dut

    Serial number  Input the new value of the serial number and confirm the input.  The new value is adopted and you will be returned to the previous window. - 34 - © SIKA • Ba_TP37-TP3M_en • 02/2018...
  • Page 35: Test Task

    TP37 / TP3M Operation 5.6 Test Task Here, you can select and manage existing test tasks. You can create new test tasks or tailor unprotected test tasks to your requirements. From the parameter field "Test Task" of the main window, you get to the window "Select Test Task".
  • Page 36: Name

    Caution! Changes to existing functions always affect all the test tasks that are linked to the function ( § 5.4). - 36 - © SIKA • Ba_TP37-TP3M_en • 02/2018...
  • Page 37: Test Specimen (Dut)

    TP37 / TP3M Operation 5.6.4 Test specimen (DUT) Here, you can link the desired test specimen with the test task. Details of the selected test specimen are displayed in the information box in the window.  Select the desired test specimen by tapping on in the selection list.
  • Page 38  Tap on the parameter field "Dwell time", input the new values and confirm the input.  The new value is adopted and you will be returned to the previous window. - 38 - © SIKA • Ba_TP37-TP3M_en • 02/2018...
  • Page 39: Alarm Settings

    TP37 / TP3M Operation Set gradient:  Tap on the parameter field "gradient", input the new values and confirm the input.  The new value is adopted and you will be returned to the previous window. Behavior at test end / Safety temperature value Using the parameter fields of the new window, you can set the safety temperature value and define the behavior at test end.
  • Page 40: Calibrator Settings

     Insert one plug of a suitable network cable in the socket of the calibrator.  Insert the other plug of the network cable in a free socket of your network. - 40 - © SIKA • Ba_TP37-TP3M_en • 02/2018...
  • Page 41 TP37 / TP3M Operation Now configure the calibrator for the network.  First set the network mode on. "DHCP" = automatic assignment of the IP address. "Manual" = Manual input of the IP address and subnet mask.  Select the desired mode by tapping in the selection list. ...
  • Page 42 Software update: For a software update, you need a new software version. The newest software version for your calibrator can be found on our homepage at "www.sika.net".  Compare the version level of your calibrator ( "System information") with the version level of the latest software.
  • Page 43 Import user files: You can use this parameter field to load previously saved user settings to the calibra- tor. Moreover, you can copy additional SIKA factory settings of measurement tasks, test specimens and functions.  Insert the USB flash drive in the calibrator and wait for 20-30 s so that the USB flash drive is recognised.
  • Page 44: Measurement Reports

     The relevant measurement data are displayed in the measurement value range. The temperature axis is scaled automatically.  Tap on the (+/-) symbols next to the measurement value range to change the scaling of the time axis. - 44 - © SIKA • Ba_TP37-TP3M_en • 02/2018...
  • Page 45: Testing Process / Calibration

    TP37 / TP3M Testing process / Calibration Testing process / Calibration CAUTION! Risk of burns! The calibrator can become very hot when in operation. Touching hot parts can result in serious injuries.  Never touch the metal block, the tank, the adapter sleeve or the test specimen at temperatures above 35 °C or below 10 °C.
  • Page 46 After the testing process:  Allow the calibrator to cool to room temperature.  Note the instructions in the section "Test tasks" ( § 3.3).  Clean the calibrator ( § 8.2). - 46 - © SIKA • Ba_TP37-TP3M_en • 02/2018...
  • Page 47: Problems

    Contact your supplier or directly to SIKA. Please send the device for repair with a brief de- scription of the problem, the environmental conditions and the length of time the device was operational before the problem occurred.
  • Page 48: Maintenance And Cleaning

    Maintenance and cleaning TP37 / TP3M The following measures must be taken before you send the unit to SIKA for repair:  Clean the device thoroughly. This is of extreme importance if the medium is hazardous to health, i.e. caustic, toxic, carcinogenic or radioactive etc.
  • Page 49: Cleaning

    TP37 / TP3M Maintenance and cleaning Recalibration:  Send the calibrator to SIKA for Recalibration ( § 8.3) after 36 months or after a maxi- mum of 500 operating hours. Calibration liquid: Calibration liquids get fouled or age with time. This depends to a great extent on the type of liquid and the usage behaviour.
  • Page 50 IMPORTANT! Temperature fuse at over-temperature! An air flow that is too low can result in the temperature fuse getting triggered. The cali- brator is then rendered operationally disabled, and must be sent to SIKA.  Ensure that the grille openings are always clear.
  • Page 51: Recalibration

    The calibrator should, depending on the application situation, be inspected at appropriate in- tervals on the basis of DIN ISO 10012. We recommend you to return the calibrator to SIKA at intervals of max. 36 months or approx. 500 operating hours for recalibration and readjust- ment.
  • Page 52: Decommissioning And Disposal

     Take the TP37 / TP3M to your local recycling plant  send the TP37 / TP3M back to your supplier or to SIKA. * WEEE reg. no.: DE 25976360 - 52 - © SIKA • Ba_TP37-TP3M_en • 02/2018...
  • Page 53: Technical Data

    TP37 / TP3M Technical data 10 Technical data The technical data of customised versions may differ from the data in these instructions. Please observe the information specified on the type plate. 10.1 Characteristics TP37 / TP3M Type Characteristics 37200E.2 37165E.2 37166E.2 37700E 3M165E.2...
  • Page 54 < 80% to 31 °C, reducing linearly up to 50% at 40 °C (non-condensing environment) Operating conditions: - Location • Altitude Interiors • up to 2000 m - Operating position Standing upright/vertically Application area tested according to EN 61326-1, class A (industrial sector) - 54 - © SIKA • Ba_TP37-TP3M_en • 02/2018...
  • Page 55: Heating Up And Cooling Times

    The following times are guide values for a room temperature of 23 °C without settling processes. 10.2.1 TP 37200E.2 • TP 37165E.2 • TP 37166E.2 • TP 37700E (dry block) 10.2.2 TP 37166E.2 (Infrared and surface) Technical changes reserved...
  • Page 56: Tp 3M165E.2

    Technical data TP37 / TP3M 10.2.3 TP 3M165E.2 Dry block, infrared and surface Micro bath: Water and silicone oil - 56 - © SIKA • Ba_TP37-TP3M_en • 02/2018...
  • Page 57: Tp 3M255E

    TP37 / TP3M Technical data 10.2.4 TP 3M255E Dry block, infrared and surface Micro bath: Water and silicone oil Technical changes reserved - 57 -...
  • Page 58: Guideline Dakks-Dkd-R 5-4

    If control measurements with a calibrated ther- mometer are not taken annually, recalibration of the temperature block calibrator is strongly rec- ommended. © Deutsche Akkreditierungsstelle GmbH (German Accreditation Body) • E-Mail: contact@dakks.de • Internet: www.dakks.de - 58 - © SIKA • Ba_TP37-TP3M_en • 02/2018...
  • Page 59 TP37 / TP3M For your notes Technical changes reserved - 59 -...
  • Page 60 TP37 / TP3M Sensors and Measuring Instruments Flow Measuring Instruments Test and Calibration Instruments SIKA Dr. Siebert & Kühn GmbH & Co. KG Struthweg 7–9 D-34260 Kaufungen  Germany  +49 (0)5605 803-0  +49 (0)5605 803-54 info@sika.net  www.sika.net - 60 - ©...