SIKA TP 38 Series Operating Instructions Manual

SIKA TP 38 Series Operating Instructions Manual

Precision temperature calibrator

Advertisement

Operating instructions
Precision Temperature Calibrator
TP 38 ... / TP 38 ... E

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TP 38 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for SIKA TP 38 Series

  • Page 1 Operating instructions Precision Temperature Calibrator TP 38 … / TP 38 … E...
  • Page 2: Table Of Contents

    13.4 Technical data of the sensor inputs ....................43 Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 3: Instrument Description And Intended Use

    TP 38 165 E and TP 38 650 are supplied without an integrated measuring instrument. Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 4: Safety Instructions

    Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 5: Ground Wire Monitoring (Tp 38 650 And Tp 38 650 E)

    : 05605-803 0   05605-803 54 info@sika.net  www.sika.net Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 6: Unpacking And Inspecting The Delivery

    Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 7: The Control Panel

    Control / Calibrate switch Clear keys for deleting data Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 8: The Metal Calibration Block

    The external resistance thermometer is connected using a 7-pole connector. Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 9: Pc Connection, Power Switch And Type Label

    Pin 2) RX Pin 3) TX Pin 5) GND Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 10: Components Of The Calibrator Bottom

    (e.g. chapter 2, chapter 4) of the operating instructions. Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 11: Accessories (Included In The Scope Of Delivery)

    A data cable for connecting the calibrator to the RS 232 interface can be ordered from SIKA. Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 12: Dkd Calibration Certificate

    Without verification measurements using a calibrated thermometer, we strongly recommend annual recalibration of the dry well calibrator. Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 13: Commissioning Of The Calibrator

    Switch on the calibrator at the power switch. Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 14: Start Procedure With Tp 38 650 / Tp 38 650 E

    = 120 °C at a dwell time of t = 15 min. Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 15: Operation Of The Calibrator

    Select mA Check switch Select switch DataLogger Select Logger Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 16 (please also refer to chapter 6.1.4) Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 17: Starting The Calibrator

    The individual serial number of the calibrators is specified in the Info menu and additionally on the type label on the rear of the calibrator. Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 18: Description Of The Control Mode

    Ref displays the value of the external reference. Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 19 Press the Control key once to start the controller. Calibration mode will be called. Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 20 Fig. 5a: Display with external reference Fig. 5b: Display without external reference Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 21 When the internal measuring instrument is switched off and on again, all the probes are switched to "OFF". Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 22: Settings And Descriptions In Select Mode

    *Since the step test is the basis of DataLogger it is first described. Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 23 The "4 mm terminals"-connection does not include setting options since resistance thermometers can only be connected to 4 mm terminals. Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 24 Below the "Connection” line you have to select, whether the probe is connected to a 4 mm terminals or a TC- plug. Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 25 The "4 mm-terminals” connection does not include setting options since mA power signals can only be connected to 4 mm terminal. Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 26 Here you can define the gradient (Kelvin per minute), with which the calibrator passes through the selected temperature range. Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 27 The "4 mm-terminals” connection does not include setting options since temperature switches can only be connected to 4 mm-terminals. Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 28 Length of the dwell time on the programmed temperature level  Cycles Number of repetitions Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 29 (mA) as the probe Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 30 Select a test specimen (cap. 6.5.1 until 6.5.3). Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 31 TP 38… of SIKA you can continue the process trouble-free (e. g. construction of calibration tickets). Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 32 The selection is displayed in Control mode. Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 33 Control mode and can be reactivated via the Control key. Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 34 Fig. 18: Submenu select Info CONTROL OFF 13:24:56 Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 35: Description Of The Switch Test Mode

    The switch is open Fig. 19: Switch test mode Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 36: Description Of The Step Test Mode

    Fig. 20: Step test mode Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 37: Cooling Of The Block And Replacing The Insert

    If a replaced fuse blows repeatedly, the calibrator is probably defective. In this case, send the calibrator to the manufacturer for repair. Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 38: Eliminating Faults

    Fax no.: (+49) 5605 - 803 - 54 eMail: info@sika.net Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 39: Transport And Storage

    TP 38… / TP 38… E series to us carriage prepaid. Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 40: Technical Data

    0...90 % relative humidity Protection class IP20 CE conformity EN61326-1, EN61010-1 Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 41: Tp 38 650 / Tp 38 650 E (Ambient Temperature Up To 650 °C)

    0...90 % relative humidity Protection class IP20 CE conformity EN61326-1, EN61010-1 Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 42: Heat-Up And Cool-Down Times

    3:02 min 40 °C to 25 °C 1:09:36 min Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...
  • Page 43: Technical Data Of The Sensor Inputs

    DIN TC- connection Pt100 DIN class B (+/- 0.3°C + 0.005 * [ T ]) Subject to technical changes Ba_TP38-E_en 12/2014 SIKA Dr.Siebert & Kühn GmbH & Co.KG  Struthweg 7–9  D-34260 Kaufungen   +49 5605 803-0  +49 5605 803-54  info@sika.net  www.SIKA.net...

Table of Contents