Advertisement

Moving
Rehabilitation
Forward™
Electromagnetic
Compatibility (EMC)
Tables
User Manual
Operation & Installation
Instructions for:
Intelect Shortwave
Optional Equipment:
020-453218 - Capacitive Plate Electrodes
020-453216 - Capacitive Plate Electrodes
020-453267 - Soft Rubber Electrode
020-453267 - Soft Rubber Electrode
020-969553 - Diplode with Cable (Coil Field Electrode)
020-453263 - Diplode (Coil Field Electrode)
3813- 115V
3814- 230V

Advertisement

Table of Contents
loading

Summary of Contents for Chattanooga Intelect Shortwave 400 Series

  • Page 1 Moving Rehabilitation Forward™ User Manual Operation & Installation Instructions for: Intelect Shortwave 3813- 115V 3814- 230V Optional Equipment: 020-453218 - Capacitive Plate Electrodes 020-453216 - Capacitive Plate Electrodes 020-453267 - Soft Rubber Electrode 020-453267 - Soft Rubber Electrode 020-969553 - Diplode with Cable (Coil Field Electrode) 020-453263 - Diplode (Coil Field Electrode) Electromagnetic Compatibility (EMC)
  • Page 2: Table Of Contents

    TABLE OF CONTENTS Intelect® Shortwave 400 FOREWORD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Patient Side .
  • Page 3 TABLE OF CONTENTS Intelect® Shortwave 400 Switching on the unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 Repairs .
  • Page 4: Foreword

    FOrEwOrd Intelect® Shortwave 400 This manual has been written for the owners of the Intelect® Shortwave 400 system. It contains general information on the operation, precautionary practices, and maintenance information. In order to maximize use, efficiency, and the life of your unit, please read this manual thoroughly and become familiar with the controls, as well as the accessories, before operating the unit.
  • Page 5: Safety Instructions

    SAFETy iNSTruCTiONS Intelect® Shortwave 400 GENErAL iNFOrmATiON The Intelect Shortwave 400 therapy unit and the accompanying components and individual elements fulfill individually and as a unit, the currently valid safety standards and comply with the stipulations of IEC 601 and the medical products regulations. The unit and its external components (accessory elements) are safe if used properly and in compliance with the explanations and instructions provided in this documentation.
  • Page 6: Precautionary Definitions

    SAFETy iNSTruCTiONS Intelect® Shortwave 400 PrECAuTiONAry dEFiNiTiONS The precautionary instructions found in this section and throughout this manual are indicated by specific symbols. Understand these symbols and their definitions before operating this equipment. The definition of these symbols are as follows: Explosion Hazard CAUTION Text with an “Explosion Hazard”...
  • Page 7: Cautions

    SAFETy iNSTruCTiONS Intelect® Shortwave 400 CAUTION CAUTION • • Inspect cables and associated connectors before each use. Read, understand, and practice the precautionary operating • instructions. Know the limitations and hazards associated with using Remove hearing aids prior to treatment. any shortwave device.
  • Page 8: Warnings

    SAFETy iNSTruCTiONS Intelect® Shortwave 400 WARNING WARNING • • Since relatively high powers are used, there is the possibility of Do not store or operate the unit in a dusty environment. • producing shock, localized burns, and cataracts if the patient is Care must be taken when operating this equipment around other unaware of the heat due to reduced thermal sensation, or if the equipment.
  • Page 9: Dangers

    SAFETy iNSTruCTiONS Intelect® Shortwave 400 WARNING DANGER • Induction field electrodes that are operated without a patient could • The unit, electrodes and cables may not be sterilized using be destroyed due to overheating. steam or gas. • • In case of display failure or other obvious defects, switch the unit Never clean the unit with abrasives, disinfectants or off immediately by means of the power switch and notify your solvents that could scratch the housing or damage the...
  • Page 10 SAFETy iNSTruCTiONS Intelect® Shortwave 400 DANGER DANGER • • Never, under any circumstances, attempt to hold any of the Shortwave diathermy should not be used on patients who electrodes in your hands during therapy. have any implanted metallic lead or any implanted system •...
  • Page 11: Personal Safety

    SAFETy iNSTruCTiONS Intelect® Shortwave 400 PErSONAL SAFETy In case of improper or unauthorized use of the unit, the • The electrode connection cable is connected properly operator, the patient or other persons may be subjected to and is not cross-routed (which may cause capacitive the danger of electric shock due to high voltage produced short circuits).
  • Page 12 SAFETy iNSTruCTiONS Intelect® Shortwave 400 PErSONAL SAFETy (CONTiNuEd) * The affected parts of the body should be unclothed Before using the unit, speak with the patient to make during treatment, since accumulation of moisture on the sure that he is in a comfortable position during the entire skin or in folds can cause local overheating of the skin.
  • Page 13: Protection Of The Unit

    SAFETy iNSTruCTiONS Intelect® Shortwave 400 PrOTECTiON OF ThE uNiT Before connecting the unit, make sure that: WARNING • The voltage rating on the safety label corresponds to the available system voltage. Improper installation, operation or maintenance of the • The frequency rating on the rating plate corresponds to shortwave therapy unit may result in malfunctions of this the system frequency.
  • Page 14: Safety Instructions

    SAFETy iNSTruCTiONS Intelect® Shortwave 400 PrOTECTiON OF ThE uNiT (CONTiNuEd) • Clean and disinfect the unit only when the Mains Do not connect the unit to the Mains Power Supply until Power Supply is deactivated (power switch off, power these requirements have been fulfilled: plug disconnected).
  • Page 15: Indications

    iNdiCATiONS Intelect® Shortwave 400 General C. Achillodynia Shortwave therapy is the application of electromagnetic Definition: Irritation of the Achilles tendon energy to the body at shortwave frequencies. Shortwave Goal of treatment: Stimulation of perfusion, elimination therapy equipment normally uses 27.12 MHz to produce the desired deep heating effect on the tissues of the body.
  • Page 16 iNdiCATiONS Intelect® Shortwave 400 G. Epicondylitis K. ischialgia Definition: Tennis elbow, inflammation of tendon Definition: Pain in ending of nervus ischiadicus, always attachments on cubital or radial part of elbow joint radicular, usually caused by damaged intervertebral disc (humeral) Goal of treatment: Analgesia, tone reduction, hyperaemisation of the affected muscles Goal of treatment: Pain relief, elimination of irritation h.
  • Page 17 iNdiCATiONS Intelect® Shortwave 400 r. raynaud´s disease O. Neuralgia / neuritis Definition: Acute arterial blood supply in the fingers Definition: Nerve pain, nerve inflammation (angiospasms) While neuritis is characterized by clear symptoms of Goal of treatment: Reduction of the frequency, dysfunction, neuralgia refers to conditions where pain is intensity and duration of the attacks primary, normally during the night...
  • Page 18: Indications

    iNdiCATiONS Intelect® Shortwave 400 u. Tendovaginitis Dosage Levels according to Schliephake Dosage 1: No perceptible heat (athermal, subliminal) Definition: Inflammation of tendon and sheath; Painful grating or chafing of the affected tendon after Dosage 2: Slight warming (heat just perceptible) overstraining or dull trauma Dosage 3: Pleasantly warm (pleasant, tolerable Goal of treatment: Inflammation stop, pain relief, free...
  • Page 19: Contraindications

    CONTrAiNdiCATiONS Intelect® Shortwave 400 General • The following list of contraindications – which is by no Patients whose condition could be negatively affected by means to be regarded as being comprehensive – should heat. always be observed when applying shortwave treatment for •...
  • Page 20 CONTrAiNdiCATiONS Intelect® Shortwave 400 Contraindications (continued) • Sudeck’s syndrome, stage I and II • Implants, areas where implants have been removed, damaged implants or metal inclusions² • Basedow’s disease (irradiation could cause serious states of agitation) • Implants that could be impaired by shortwave irradiation • Varicose veins (irradiation could cause congestive pain) • Swellings that still feel warm • Thermohyperesthesia...
  • Page 21: Contraindications

    CONTrAiNdiCATiONS Intelect® Shortwave 400 Contraindications (continued) • The output power must always be set according to the subjective response of the patient. Therefore, special care • Particular care must be taken if the patient’s clothing is must be taken in case of patients with a diminished wet or damp, since the garments may heat up faster and (Dosage levels capacity for perception of heat.
  • Page 22: Installation

    iNSTALLATiON Intelect® Shortwave 400 • By selecting a suitable location for setting up the unit or Requirements for Installation by means of structural measures, contact during the Before the unit can be installed and put into operation, treatment by the personnel or the patient with certain requirements must be fulfilled in the building where conductive materials that are earthed or have a high the unit is to be operated.
  • Page 23: Unpacking The Unit

    iNSTALLATiON Intelect® Shortwave 400 Unpacking the Unit The unit is generally delivered with the packaging material DANGER supplied by the manufacturer. Since the unit weighs approximately 60 kg (132 lbs), it must be unpacked by at least 2 persons. In case of damage from transport that could endanger personal safety, the unit must not be connected to the Proceed as follows: Mains Power Supply.
  • Page 24: Nomenclature

    NOmENCLATurE Intelect® Shortwave 400 OPErATOr SidE 1. Capacitive Plate Electrode Ø 120 mm 2. Electrode connection cable with sponge rubber covering 3. Self-locking plastic electrode arm, extractable 4. Cable clamp for electrode connection cable 5. Control panel with graphic display 6.
  • Page 25: Patient Side

    NOmENCLATurE Intelect® Shortwave 400 PATiENT SidE 1. Self-locking plastic electrode arm, extractable 2. Male end of the electrode connection cable 3. Electrical Safety Protection (fuses) 4. High-load non-heating Appliance Plug (Power Cord) 5. Serial Decal 6. Female end of the electrode connection cable 7.
  • Page 26: Accessories

    NOmENCLATurE Intelect® Shortwave 400 ACCESSOriES WARNING 1. Capacitive Plate Electrode Ø 165 mm 2. Capacitive Plate Electrode Ø120 mm (Standard accessory) Keep all electrodes, accessories, and their cords separated during treatment by using the cable clips located on the 3. Capacitive Plate Electrode Ø 80 mm side of the arm extenders.
  • Page 27: Nomenclature

    NOmENCLATurE Intelect® Shortwave 400 ACCESSOriES (CONTiNuEd) Figure 4 1. Soft Rubber Electrode Soft Rubber Electrodes - 250 x 145 mm Dielectric Field 2. Soft Rubber Electrode 180 x 120 mm 3. Felt layer with linen bag for 1 and 2 4.
  • Page 28: Specifications

    SPECiFiCATiONS Intelect® Shortwave 400 uNiT SPECiFiCATiONS Standard Weight (without electrodes) ..... . 60 kg (132 lbs) Ambient temperature ..... +10 °C to 40 °C (50 °F to 104 °F) Relative Humidity .
  • Page 29: Description Of Functions

    SPECiFiCATiONS Intelect® Shortwave 400 dESCriPTiON OF FuNCTiONS Introduction The Intelect Shortwave 400 therapy unit can produce dielectric warming by means of electric or electromagnetic fields of varying intensity in essentially any region of the body and can therefore be used for a wide variety of applications. Treatments can be carried out using either the capacitive field or the inductive field method.
  • Page 30: Shortwave Therapy In The Capacitive - Dielectric Field

    SPECiFiCATiONS Intelect® Shortwave 400 dESCriPTiON OF FuNCTiONS (CONTiNuEd) Shortwave Therapy in the Capacitor (Dielectric) Field With the capacitive field method, the body part being treated is located in the high-frequency electric field between two insulated electrodes. The body and the electrodes together form a capacitor.
  • Page 31: Shortwave Therapy In The Inductive Field

    SPECiFiCATiONS Intelect® Shortwave 400 dESCriPTiON OF FuNCTiONS (CONTiNuEd) Shortwave Therapy in the Induction Field The inductive field produces the high-frequency electric currents within the body tissue by means of induction. The high-frequency circuit of closed currents occurring in the tissue as a result of induction are referred to as eddy currents. The density of these eddy currents, which is important for the warming effect, is proportional to the electrical conductivity of the tissue.
  • Page 32: Program Description

    SPECiFiCATiONS Intelect® Shortwave 400 PrOGrAm dESCriPTiON Display and Control Elements 1. Intelect Shortwave 400 operating element 2. Graphic display 3. Multi-function keys 4. Stop-key (power output shut-off ) 5. Important: Read operating instructions 6. Multi-function rotary switch with push-button function (Enter) NOTE: Device of type BF (insulated non-earthed application element)
  • Page 33: Main Menu

    SPECiFiCATiONS Intelect® Shortwave 400 PrOGrAm dESCriPTiON (CONTiNuEd) Main Menu 5. Function key for activation of the output power Main menu after unit is switched on parameter 6. Function key for activation of the treatment duration parameter 7. Function key for activation of the operating mode parameter 8.
  • Page 34 SPECiFiCATiONS Intelect® Shortwave 400 PrOGrAm dESCriPTiON (CONTiNuEd) Main Menu During Treatment 6. Function key for activation of the output power parameter 7. Function key for activation of the treatment duration parameter 8. Function key for activating the frequency parameter in pulsed mode 9.
  • Page 35: Indications Menu

    SPECiFiCATiONS Intelect® Shortwave 400 PrOGrAm dESCriPTiON (CONTiNuEd) Indications Menu (Indications Menu; Main Page) 5. Function key for confirmation of parameters of the selected disorder (Enter); alternatively, the push-button function of the multi-function rotary switch can be used 6. Function key for calling up further information (application of electrodes, comments on therapy) Figure 11 Indications Menu;...
  • Page 36 SPECiFiCATiONS Intelect® Shortwave 400 PrOGrAm dESCriPTiON (CONTiNuEd) Indications Menu (Index Page) 1. Medical field, not selected 2. Medical field, selected (inverse display) 3. Function key for confirming the selected medical field (Enter); alternatively, the push-button function of the multi-function rotary switch can be used 4.
  • Page 37 SPECiFiCATiONS Intelect® Shortwave 400 PrOGrAm dESCriPTiON (CONTiNuEd) Indications Menu (Info Page) 1. Disorder with duration of treatment and operating mode display field 2. Goal of treatment display field 3. Applicator display field * 4. Electrode-Skin Distance (ESD) display field 5. Recommended treatment dosage display field ** 6.
  • Page 38: Programming Menu

    SPECiFiCATiONS Intelect® Shortwave 400 PrOGrAm dESCriPTiON (CONTiNuEd) Programming Menu 1. Program position selected display field. 2. Free program position display field. 3. Function key for return to main menu. 4. Function key for deleting the parameters of the selected program position. 5.
  • Page 39: Basic Settings Menu

    SPECiFiCATiONS Intelect® Shortwave 400 PrOGrAm dESCriPTiON (CONTiNuEd) Basic Settings Menu 1. Pulse duration in pulsed mode display field 2. Transmitter volume display field 3. Service code display field 4. Language selection menu font display field 5. Graphic display brightness display field 6.
  • Page 40: Operation

    OPErATiON Intelect® Shortwave 400 PuTTiNG ThE uNiT iNTO OPErATiON Preparing the Unit for Operation DANGER The unit has been completely assembled in the factory and is ready for use except for connection of the The unit and the electrodes must be positioned so that electrodes and the power cord.
  • Page 41: Positioning Of The Electrodes

    OPErATiON Intelect® Shortwave 400 PuTTiNG ThE uNiT iNTO OPErATiON Positioning the Electrodes Position the required electrodes on the part of the body to Correct positioning of the electrodes for the Edge Effect be treated according to the medical indication. Refer to the allows for equal distribution of the concentrated heating indications table at the end of this instruction manual and effect.
  • Page 42 OPErATiON Intelect® Shortwave 400 PuTTiNG ThE uNiT iNTO OPErATiON (CONTiNuEd) Local overheating can also occur due to electrode Surface warming can be reduced by increasing the constrictions. This can be prevented by increasing the electrode distance. The use of the diode is recommended distance (e.g.
  • Page 43 OPErATiON Intelect® Shortwave 400 PuTTiNG ThE uNiT iNTO OPErATiON (CONTiNuEd) Below you will find several examples for correct placement of electrodes: Figure 19 Positioning of Electrodes – Normal Cases 3. Uneven irradiation by the use of different electrodes 1. Even irradiation of extremities in the transverse field 4.
  • Page 44: Setting The Electrode-Skin Distance

    OPErATiON Intelect® Shortwave 400 PuTTiNG ThE uNiT iNTO OPErATiON (CONTiNuEd) Setting the Electrode-Skin Distance The full power required for successful depth therapy is provided by the unit by using a large Electrode-Skin Distance (ESD). For treatment near the surface, in which the power must be limited in accordance with the respective therapy, a small Electrode-Skin Distance is required.
  • Page 45: Switching On The Unit

    OPErATiON Intelect® Shortwave 400 PuTTiNG ThE uNiT iNTO OPErATiON (CONTiNuEd) Setting the Electrode-Skin Distance (continued) Setting the Parameters Parameters can be set in three ways: The soft rubber electrodes adapt to the shape of the body and are held to the body with elastic bands or with small 1.
  • Page 46: Selecting Parameters In The Indications Menu

    OPErATiON Intelect® Shortwave 400 PuTTiNG ThE uNiT iNTO OPErATiON (CONTiNuEd) Selecting Parameters in the Indications Menu Set the parameter by selecting a disorder in the indications menu as follows: 1. Select the indications menu by pressing the function key (Indic.) in the main menu. 2.
  • Page 47: Basic Operations

    OPErATiON Intelect® Shortwave 400 BASiC OPErATiON (CONTiNuEd) Beginning the Treatment Treatment is always started from the basic menu by turning up the output power by means of the multi-function rotary switch from zero to the desired value. The treatment can be interrupted at any time by pressing the STOP key. DANGER In order to prevent excessive warming of tissue, the effective output power, as summarized in the...
  • Page 48: Interrupting/Ending Treatment

    OPErATiON Intelect® Shortwave 400 BASiC OPErATiON (CONTiNuEd) Beginning the Treatment (continued) Begin the treatment as follows: 1. Make sure that the electrodes are in the correct position. 2. Check the parameter setting. 3. Select the output parameter by pressing the function key (W). 4.
  • Page 49: Changing The Basic Settings

    OPErATiON Intelect® Shortwave 400 BASiC OPErATiON (CONTiNuEd) Changing the Basic Settings Switch off the power switch. Press the multi-function rotary switch and hold it down. Switch the unit on by means of the power switch and wait (with the multi-function rotary switch depressed) until the basic settings menu is displayed.
  • Page 50: Maintenance

    mAiNTENANCE Intelect® Shortwave 400 mAiNTENANCE ANd rEPAirS Routine Maintenance As the manufacturer, we are responsible for the safety and reliability of the unit only if it is used in accordance with the operating instructions. Safety inspections, maintenance, repairs and modifications may be performed only by our company or expressly authorized service personnel.
  • Page 51 mAiNTENANCE Intelect® Shortwave 400 mAiNTENANCE ANd rEPAirS (CONTiNuEd) Cleaning and Disinfection (continued) Clean and disinfect the unit and accessories on all external surfaces with a slightly damp cleaning cloth. Use a commercially available cleaning agent that is intended for use in medical facilities. Wait until the unit is completely dry before operating it again.
  • Page 52: Safety Inspection

    mAiNTENANCE Intelect® Shortwave 400 mAiNTENANCE ANd rEPAirS (CONTiNuEd) Safety Inspections The following safety inspections must be performed on this unit. This must be done by persons who, based on training, knowledge or practical experience, are capable of conducting the inspections correctly and independently. Visual Inspection (Daily) • Housing not deformed? • Power Cable undamaged?
  • Page 53: Electrical Tests

    mAiNTENANCE Intelect® Shortwave 400 mAiNTENANCE ANd rEPAirS (CONTiNuEd) Electrical Test (Acc. to IEC 601-1) (12 month intervals or as dictated by you Health Care Facility) • Earth leakage current < 0,5 mA (normal condition) • Earth leakage current < 1,0 mA (single fault condition) • Housing leakage current < 0,1 mA (normal condition) • Housing leakage current < 0,5 mA (single fault condition) • Protective conductor resistance < 100 mΩ NOTE: It is the responsibility of the Health Care Facility to verify that the unit complies with the facility, local and national Earth Leakage limits.
  • Page 54: Error Messages

    mAiNTENANCE Intelect® Shortwave 400 mAiNTENANCE ANd rEPAirS (CONTiNuEd) Error Messages (Requires authorized service personnel) Both operator errors and internal errors in the unit are displayed as an error code with text information. Operator Errors E01: Please check electrodes and cables Cause: An optimum tuning point could not be found;...
  • Page 55: Repairs

    mAiNTENANCE Intelect® Shortwave 400 mAiNTENANCE ANd rEPAirS (CONTiNuEd) Repairs DANGER This product complies with WEEE Directive 2002/96/EG (waste electrical and electronic equipment). Separate collection for electrical Do no perform unauthorized repairs under any and electronic equipment. circumstances. Service Should the unit require service, warranty, or repair, please contact the selling dealer or your local DJO customer service.
  • Page 56: Accessories

    ACCESSOriES Intelect® Shortwave 400 STANdArd ACCESSOriES NOTE: Power Cords are not listed; however the Power Ref. description Cord shipped with the unit will accommodate the electrical requirements for the country of 39-14762 Indicator Discharge Tube use. 39-020453217 Capacitive Plate Electrode 120mm diameter 28256 Intelect Shortwave 400 User Manual OPTiONAL ACCESSOriES...
  • Page 57: Warranty

    wArrANTy Intelect® Shortwave 400 DJO, LLC ("Company") warrants that the Intelect Shortwave 400 ("Product") is free of defects in material and workmanship. This warranty shall remain in effect for two years (24 months) from the date of original consumer purchase. If this Product fails to function during the two year warranty period due to a defect in material or workmanship, at the Company's Option, Company or the selling dealer will repair or replace this Product without charge within a period of thirty (30) days from the date on which the Product is returned to the Company or the dealer.
  • Page 58 DJO is an iSO 13485 certified company DJO France SAS Centre Européen de Fret 64990 Mouguerre, France T: 1-800-592-7329 USA T: + 1-317-406-2209 F: + 1-317-406-2014 chattgroup.com © 2012 DJO, LLC. All rights reserved. 28256_C...

This manual is also suitable for:

Intelect shortwave 400 3814Intelect shortwave 400 3813

Table of Contents