Measurement of average torque
Objective
Simple monitoring of torque for normal operations
Measurement procedure
Makes possible measurement of the minimum,
average and maximum values among sets of mea-
surements, each set consisting of up to ten measure-
ments.
1. The measurement results are automatically num-
bered, from 0001 to 0010. Negative displays are
not recorded. Measurements in average mode are
automatically deleted when the meter exits aver-
age mode.
2. If less than 10 measurements are made, the dis-
play shows only those measurements.
0001
0002
0003
1st measurement deleted
0001
0002
0003
0001
0002
0003
Switch functions in average measurement mode
Switches used
  ③ 
RESET
AVE
③ 【 RESET 】 switch
Puts the instrument into a state of readiness to
begin measurement when in average measure-
ment mode. Press to end measurement or dis-
play of measurement sets.
⑥ 【 AVE 】 switch
Press for 2 seconds to enter or exit average
measurement mode. The LED lights when in
average measurement mode. The figure show
the average measurement and display cycles.
Example of average measurement
(Average measurement mode, LED on.)
Ready for measurement
Use Fidaptor or other adapter to fit torque
Torque measurement
tool into the torque meter socket.
Press once. "0001" flickers. (Same for auto-reset.)
③【RESET】 switch
Repeat measurements until sufficient data is
obtained.
Display of average measurement data
(Data is displayed while in average measurement mode and is deleted
when the mode is terminated.)
⑥  Each time the 【AVE】 switch is pressed minimum, average and maxi-
mum values are displayed flickering.
Fig. 1
Min
0010
0009
0010
0009
0010
11th measurement recorded
Ave
Max
Notes
1. Be sure to set 【 MODE 】 switch to "PEAK" or
"F.PEAK" before entering average measurement mode.
2. The 【 UNIT 】 switch does not function in average
measurement mode.
■ Changing the Fidaptor's shock-
absorbing spring (Model HP-100)
Two springs are supplied with Model HP-100. The
black spring is high-tension and the yellow spring
is low-tension. The 5 mm Fidaptor is shipped with
the black spring attached. For low-torque measure-
ment, replace with the yellow spring.
 Fit the driver bit into the head of the Fidaptor.
Turn the driver on in the reverse direction to
loosen the threaded shaft and change the springs.

■ Care of the Fidaptor

Please read carefully the following points about
handling the Fidaptor.
 Using the Fidaptor
1. Never place a load on the Fidaptor that exceeds
the indicated measurement capacity.
2. Use the Fidaptor that is designed for the torque
level of the tool to be measured.
3. Keep the Fidaptor loose, and after measurement
remove it from the socket.
4. Keep the driver and Fidaptor upright during mea-
surement. Keep torque from the upper surface to
within 5 kg, or 2 kg, in the case of a low-torque
Fidaptor.
5. When conducting a series of measurements, ap-
ply grease* to parts ① , ② and ⑤ of the Fidap-
tor, as shown in the fi gure on page 11.
- 10 -