Counter Subsystem - Hantek HDG 2000 Series User Manual

Arbitrary waveform generator
Table of Contents

Advertisement

This setting is non-volatile; it will not be changed by power cycling or *RST.
CALibration:STORe
Takes calibration constants in volatile memory (CALibration?), and places them in nonvolatile
memory, where they will not be changed by power cycle.
Parameter
none
Store calibration constants into non-volatile memory: CAL:STOR
CALibration:VALue <value>
Specifies the value of the known calibration signal.
Parameter
Numeric, default 0
-
This setting is non-volatile; it will not be changed by power cycling or *RST.

5.2.6 COUNter Subsystem

COUNter ON|OFF
COUNter:MEASure?
COUNter:GATEtime
COUNter:GATEtime?
COUNter:TYPE
COUNter ON|OFF
Enable or disable frequency meter.
Parameter
ON|OFF, default OFF
COUNter ON
COUNter:MEASure?
Get current counter measurement value.
Parameter
COUNter:GATEtime AUTO|USER1|USER2|USER3|USER4|USER5|USER6
COUNter:GATEtime?
Set GATEtime
HDG2000 Series Arbitrary Waveform Generator
Typical Return
Typical Return
none
Typical Return
ON , or OFF
Typical Return
Current frequency or counter
SCPI Programming Reference