Parameter
ON|OFF, default OFF
Enable burst mode: SOURce1:BURSt ON
Output phase is set to 0 when burst is enabled.
To avoid multiple waveform changes, enable the burst mode after configuring the other burst
parameters.
The instrument will not allow the burst mode to be enabled at the same time that sweep or any
modulation mode is enabled. When you enable burst, the sweep or modulation mode is turned
off.
SOURce<n>:BURSt:TRIGger:SLOPe POSitive|NEGative
SOURce<n>:BURSt:TRIGger:SLOPe?
Select POSitive or NEGative of the external trigger signal to start sweep output.
Parameter
POSitive|NEGative, default POSitive
Select NEGative of the external trigger signal to start sweep output:
SOURce<n>:BURSt:TRIGger:SLOPe NEGative

5.2.5 CALibration Subsystem

The CALibration subsystem is used to calibrate the instrument.
CALibration:SETup <step>
CALibration?
CALibration:VALue <value>
CALibration:STORe
CALibration?
Performs a calibration using the calibration value (CALibration:VALue).
Parameter
none
Calibrate using the current value: CALibration?
CALibration:SETup
Modifies the volatile version of the calibration constants. Use
constants in nonvolatile memory at end of calibration.
CALibration:SETup <step>
Configures the calibration step to be performed.
Parameter
Whole number, 1 to 8
Prepare for calibration step 5: CALibration:SETup 5
HDG2000 Series Arbitrary Waveform Generator
should always precede the CALibration? query.
Typical Return
ON or OFF
Typical Return
POSitive or NEGative
Typical Return
0( pass) or +1( fail)
CALibration:STORe
Typical Return
none