Hach 910 Instrument Manual page 12

Sigma flow meter
Table of Contents

Advertisement

Specifications
Model 920
Dimensions
Weight
Enclosure Material
Enclosure Rating
Operating Temperature Range
Storage Temperature Range
Power Source
Battery Life
User Interface
Monitoring Intervals
Program Memory
Time-Based Accuracy
Units of Measurement
Data Storage
Communications
Modem
Local Terminal
Velocity Measurement Accuracy
Page 10
Specifications
16.8 cm dia. x 44.7 cm L (6.625" dia. x 17.625" L)
7.5 kg (16.5 lbs.) with batteries
PVC
NEMA 6P (IP67)
-18 to 60 °C (0 to 140 °F)
-40 to 60 °C (-40 to 140 °F)
Two Energizer EN-529, Alkaline, 6 V dc batteries
90 days typical, with a 15-minute recording interval, one level and one velocity, one
data download per week, at 10 °C (50 °F) (also affected by site conditions)
PC
1, 2, 3, 5, 6, 10, 12, 15, 20, 30, 60 minutes
Non-volatile, programmable flash, can be updated via RS232 port
±1 second per day
Level: in., m, cm, ft.
Flow: gps, gpm, gph, lps, lpm, lph, mgd, afd, cfs, cfm, cfh, cfd, m
Totalized Flow: gal., ft.³, acre-ft., L, m³
Capacity: 240 days of two level readings and two velocity readings at a 15-minute
recording intervals
Data Types: Level, velocity, and rainfall
Storage Mode: Wrap or slate
RAM Memory: 128K
Serial connection to IBM compatible computer with Hach analysis software
14400, V.32 bis,V.42, MNP2-4 error correction. V.42 bis, MNP5 data compression.
MNP10EC Cellular Protocol
RS232 at 19.2k baud
Method: Doppler ultrasonic
Transducer Type: Twin 1 MHz piezoelectric crystals
Typical minimum depth for velocity: 2 cm (0.8 in.)
Range: -1.52 to 6.10 m/s (-5 to 20 fps)
Zero Stability: 0.015 m/ss (<0.05 fp)
Accuracy: ±2% of reading
Operating Temperature: -18 to 60 °C (0 to 140 °F)
3
3
3
3
s, m
m, m
h, m
d
Specs.fm