TROUBLESHOOTING
(Continued from previous page)
Problem
Probable Causes
Conductivity
Poor connection due to moisture between
readings erratic or
contacts in connector
drifting
Dirty conductivity element
Out of calibration or improperly calibrated
Temperature compensation turned off
Loose or broken wires in connector
Damaged cable, cracked or fraying
Moisture in the unit
Damaged transmitter
Seametrics • 253.872.0284
Things to try...
Dry thoroughly. Be sure desiccant is fresh (see
Maintenance section).
Clean element
Recalibrate (see Settings and Calibration
section)
In Settings and Calibration section:
If using Non-Linear temperature
compensation, be sure Non-Linear box is
checked
If using normal temperature
compensation, be sure that the Temp.
Coef. is not set to zero. (2.1 works for
a wide variety of fluids). If set to zero,
the conductivity readings will vary with
temperature.
Repair or return for evaluation and repair
Replace cable
Return for evaluation and repair
Return for evaluation and repair