INSTRUCTION MANUAL AND REPLACEMENT PARTS
DESCRIPTION
T20 and T21 units are simple, reliable
designed for top mounting to tanks or vessels. T20 units
utilize a single switch mechanism and float. T21 tandem
units utilize two switch mechanisms and two separate
floats when widely spaced actuating levels are required.
T20 and T21 tandem models are available for any type of
open or closed vessel with either threaded or flanged type
mounting and actuating depths up to 1219 mm (48 inches)
OPERATING PRINCIPLE
A permanent magnet
is attached to a pivoted switch
actuator
. As the
float/ displacer
uid level, it raises the attraction sleeve
the magnet, which then snaps against the non-magnetic
enclosing tube
, actuating the switch. The enclosing tube
provides a static pressure boundary between the switch
mechanism and the process. On a falling level, an inconel
spring retracts the magnet, deactivating the switch.
pivot
return spring
rising level
UNPACKING
Unpack the instrument
carefully. Inspect all units for dam-
age. Report any concealed damage to carrier within 24
hours. Check the contents of the packing slip and purchase
order. Check and record the serial number for future refer-
ence when ordering parts.
®
float switches
rises following the
liq-
into the
field of
falling level
T20 & T21
Top mounting
Liquid level switches
AGENCY APPROVALS
Agency
Approval
ATEX
II 2G EEx d II C T6, explosion proof
II 1G EEx ia II C T6, intrinsically safe
CENELEC
EEx d
II C T6, explosion proof
CCE
R1 (1) 136/MI/433, explosion proof
FM
Class
I, Div. 1,
Groups C & D
Class
II, Div. 1,
Groups E, F & G, Type NEMA 7/9
FM/CSA
Non-Hazardous area
Explosion proof area –
Groups B, C, D, E, F & G Type NEMA 4X/7/9
Explosion proof area
SAA
LRS
Lloyds Register of Shipment (marine applications)
GOST/
Russian Authorisation Standards
GOSGORTECHNADZOR
Other approvals are available, consult factory for more details
➀ For CCE approved units, use the
numbers.
➁ Consult factory for proper model numbers.
ATEX explosion proof model

Advertisement

Table of Contents
loading

Summary of Contents for Magnetrol T20

  • Page 1 floats when widely spaced actuating levels are required. T20 and T21 tandem models are available for any type of open or closed vessel with either threaded or flanged type mounting and actuating depths up to 1219 mm (48 inches) OPERATING PRINCIPLE ①...
  • Page 2 Float sizes ø 76 x 127 mm (3" x 5") ø 102 mm (4") ø 114 mm (4 ") Threaded NPT connection - for T20 models only 1" ANSI Flanges - for all models 150 lbs RF 300 lbs RF...
  • Page 3 MODEL IDENTIFICATION (cont.) Select electric switch mechanism & enclosure for models T20 (see page 3 for switch ratings) All models with material code A All models with material codes B and D ATEX (IP 66) FM (IP 66) ATEX (IP 66)
  • Page 4: Installation

    INSTALLATION MOUNTING OBSERVE ALL APPLICABLE ELECTRICAL CODES AND PROPER WIRING PROCEDURES Before assembling control to tank or vessel, check thread- ed or flanged mounting nozzle for the following: NEMA 4x – Nozzle length and inside diameter must be sized correct- ly to allow for switch actuation at design levels within the maximum differential available (see table on page 4).
  • Page 5: Switch Differential Adjustment

    SWITCH DIFFERENTIAL ADJUSTMENT The standard differential of T20 and T21 Liquid Level 5a. Perform system shut-down procedures as required to Switches may be field adjusted. Adjustment may be nec- relieve pressure from tank or vessel and drain off liquid essary if a wider differential needs to be set to overcome head, if required.
  • Page 6: Troubleshooting

    “Preventive Maintenance” section. If NOTE: When in doubt about the condition or performance switch is damaged, replace it immediately. of a Magnetrol control, return it to the factory. See “Our 8. If switch mechanism is operating satisfactorily, proceed Service Policy” on back page.
  • Page 7: Preventive Maintenance

    PREVENTIVE MAINTENANCE Periodic inspections are a necessary means to keep your Magnetrol level control in good working order. This control is, in reality, a safety device to protect the valuable equipment it serves. Therefore, a systematic program of "preven- tive maintenance" should be implemented when control is placed into service. If the following sections on "what to do"...
  • Page 8 Distance to actuating level actuating level min 102 (4) min 102 (4) Actuating Actuating level level Model T20 with 1" NPT Model T20 with flange 46 (1.81) Ref. Distance to upper actuating level min 102 (4) Actuating level Distance to...
  • Page 9 FLOAT SELECTION AND INSERTION DEPTH PRESSURE RATINGS T20/T21 switches are fabricated to meet customer specific Float ratings are the maximum allowable pressure rating, insertion depth, from mounting fitting to actuating level. The even though the tank connections may have higher ratings.
  • Page 10 032-6344-002 032-6344-001 032-6344-002 032-6344-001 NEMA 4X, NEMA 7/9 Pneumatic housing 032-6302-031 032-6302-036 032-6302-033 032-6302-037 (only for T20) ➂ Mounting Flange See Replacement Mounting Flange Chart ➃ Float Guide Cage (optional) Guide Guide Cages are special order items. NOT REQUIRED Cage...
  • Page 11 REPLACEMENT PARTS cont.
  • Page 12: Service Policy

    Owners of Magnetrol products may request the return of a control; or, any part of a control for complete rebuilding or replacement. They will be rebuilt or replaced promptly. Magnetrol International will repair or replace the control, at no cost to the purchaser, (or owner) other than transportation cost if: a.