O
PERATION
How to Operate the Fluid Warmer
Before you begin
Procedure
24
The warmer must be connected to electric power and turned on.
WARNING:
Do not operate the warmer in the presence of flammable
anesthetic mixtures (with air or with oxygen or nitrous oxide).
WARNING:
Verify the fluid temperature prior to using the fluid. Refer to the
fluid manufacturer's label for recommended warming
procedures.
Do not use any fluids that are warmed above the suggested
temperature.
Do not overload the cavity.
NOTICE
Refer to topic Specification Information for the storage cavity
capacity.
To operate the warmer, do the following.
Step
Action
1.
Select
the operation mode.
Press
the irrigation button
Press
the injection button
The selected operation mode will illuminate
NOTE:
from a higher temperature to a lower temperature.
2.
Set
the temperature using the arrow buttons
NOTE:
90°F – 150°F (32°C – 66°C).
The injection temperature set-point range is 90°F – 104°F
(32°C – 40°C).
C om bi n a t i on B la nk et / F l u i d W a rm er
Personal injury hazard.
Personal injury hazard.
for irrigation mode.
for injection mode.
Open the door to cool the cavity before changing
The irrigation temperature set-point range is
O per ator 's Manu al
(Top Cavity)
.
.
Continued on next page
MN-3993 0