Maintenance Schedule
Daily
Monthly
Yearly
C om bi n a t i on B l a n ket / F lu i d W a r mer
For daily maintenance, do the following.
Check:
the air vents in the airflow insert panels are not obstructed (if applicable).
▪
all fan guards are clear and not obstructed (if applicable).
▪
the number of bottles, bags, or blankets do not exceed the maximum capacity
▪
per shelf or basket.
Clean:
any spills with a clean, lint free cloth.
▪
For monthly maintenance, do the following.
Check:
the door gasket for tears and holes. Make sure that it is firmly attached to the
▪
door. Check the seal when the door is closed.
the guards around the air temperature sensor are in place and fully secure to the
▪
warmer.
the hardware securing the warmer(s) to the wall, if applicable.
▪
Clean:
the outside of the warmer with stainless steel cleaner.
▪
vacuum (if applicable):
▪
fan openings
□
fan sail switch
□
vent openings
□
For yearly maintenance, do the following.
Check:
the set-point temperature compared to the actual temperature displayed.
▪
Check the cavity air temperature with a thermocouple placed 1" (25mm) from
□
the cavity sensor. Do not allow the thermocouple to touch any surface.
Monitor the temperature for approximately one hour in an empty cavity.
the condition of the plug and cord and replace if damaged.
▪
the controller screen for excessive wear. Make sure the controller screen displays
▪
and operates properly.
Clean:
the shelves and interior of the warmer.
▪
O pe ra t o r ' s M a n ua l
MN-39 930
▪
▪
Re v 1
3/19
▪
▪