How To Replace A Fuse - Pedigo P-2147 Operator's Manual

Combination blanket/fluid warmer
Hide thumbs Also See for P-2147:
Table of Contents

Advertisement

How to Replace a Fuse

Procedure
C om bi n a t i on B l a n ket / F lu i d W a r mer
WARNING:
Use only UL listed 20A, 250V fast acting fuses, 6.3mm x 32mm
(F1, F2). Access should be made by qualified service technicians
only.
To replace a fuse, do the following.
Step
Action
1.
Press and hold
2.
Set
the power switch
a
3.
Unplug
the power cord
inlet
on the warmer.
c
b
4.
Unscrew
the cap on the fuse holder
5.
Remove
the fuse
Ope ra tor's Ma nu a l
MN-39 930
Fire and electrical shock hazard.
the ON/OFF button
until the screen turns off.
to the OFF (0) position.
a
from the electric power source and the power
b
c
e
.
d
from the fuse holder.
e
Re v 1
3/19
T
ROUBLESHOOTING
EC-TS-004628
d
EC-TS-004675
Continued on next page
41

Advertisement

Table of Contents
loading

Table of Contents