Overload Protection; Maintenance Of Motor; Stand; Assembly Of H Stand - Scilogex D-500 User Instruction Manual

Homogenizer
Hide thumbs Also See for D-500:
Table of Contents

Advertisement

Overload Protection

When the motor overloads and the power supply cuts off, switch off the main switch to
O, disconnect the aggregate and re-start the instrument as fast as possible, without the
load in order to accelerate cooling by the built-in fan.

Maintenance of Motor

The motor does not need any maintenance and there are no parts inside which can be
repaired by the user. The only parts which are excluded from this are the carbon
brushes. Please contact or your authorized supplier for replacement and use original
spare parts only. The carbon brushes can be replaced after disconnecting the power
supply!

Stand

The D-500 homogenizer is normally delivered with either the H-600 stand or the PT 800
plate stand. It is always important to safely fix the unit even when not using the original
stand for it, as the center of gravity is relatively high.
Make sure that the motor drive is fixed safely and cannot move downwards at any time
as this may cause damage to the instrument, the user or the sample.

Assembly of H stand

This H stand consisting of a double rod, two boss heads, a safety ring and a strap clamp
to secure the vessel.
The unit comes semi-assembled as the two double rods are connected and the two
boss heads and the safety ring is fixed in place already.
Put together the foot of the stand as
following: put each of the two bolts into its
place in the middle foot piece and screw
the both side pieces together with the
provided larger screws and the allen key.

Advertisement

Table of Contents
loading

Table of Contents