Proper Use
The voltage on the name plate must match the main voltage. If it does not, do not
operate the instrument.
Screw the support rod into the flange of the drive and tighten it. Make sure that the
"nylon" washer is between the flange and the nut to avoid scratching and rubbing
marks on the flange.
Make sure that the unit is switched off when connecting or disconnecting the
dispersing shaft into the drive.
For correct use the D-500 drive must be secured to a stand with a boss head. We
recommend that for this purpose you only use the H-600 plate stand or the PT-plate
stands. For safety reasons the container should be fixed with a strap clamp (which is a
part of the H-600 stand).
Insert the dispersing element into the drive flange by pressing upwards and turning
clockwise until it locks into place. A click confirms the correct insertion. Once inserted
the dispersing element should not move any longer.
To remove the dispersing element from the flange, hold the dispersing element and
turn anticlockwise and gently remove the dispersing element downwards. The distance
between the bottom of the vessel and the shaft should not be less than 8mm.