Product Description

The D-500 Disperser is a high speed dispersing instrument. It is based on the
Rotor/Stator Technology. The shaft and rotor/stator can be disassembled for easy
cleaning. You will receive the dispersing tools completely assembled for immediate use
with your disperser. Should you dismantle these (e.g. for cleaning) please refer to page
6. This product is for laboratory use only and is used in chemical industry, cosmetic
industry, pharmaceutical industry, paint industry as well as in universities and a wide
range of medical laboratories. The unit is a high performance dispersing/homogenizing
unit and also can be used as a high speed mixing unit with certain dispersing shafts.
The drive unit can be used, based on the dispersing shaft used for volumes from 10ml
to 8,000ml. There is a wide range of dispersing shafts with diameters of 20mm, 30mm
and 40mm stator diameters. It is to be mounted on a stand and not used as a hand
held instrument.
precautions are necessary to reduce the risk of fire, electric shock and personal
injury.
• Only use the instrument for its intended purpose.
• Ensure that the correct electric voltage of the instrument and the power supply
correspond correctly.
• Do not use this instrument in a hazardous area or manner. When handling
hazardous chemicals, use appropriate hand and eye protection.
• Do not immerse electrical equipment in water.
• The drive must not be used in highly combustible areas and operated with
easily inflammable liquids. It is recommended to run the units in fume hoods
during operation.
• To avoid electrical shock, do not open housing. Remove cord from the power
source when it is being checked or serviced. This instrument should only be
opened by a qualified service personnel only.
• The D-500 is designed for continuous operation, however normally the
ultimate fineness will be reached within a few minutes. Any further dispersing
will only introduce unnecessary heat into the medium.
• The D-500 must never run without liquid – the lower slide bearing is cooled and
lubricated by the liquid phase of the treated medium. Any dry running will
destroy the slide bearing!
• Ensure that the dispersing shafts are cleaned properly after every use. When
cleaning, remove the power cord from the power source.
• Never touch the spinning rotor, or shaft, or motor side coupling parts.