Technical Specifications; Environmental Conditions For Storage And Transport; Technical Information; Part 852 Biomonitor (Ultium Biomonitor Smartlead) - Noraxon Ultium EMG BioMonitor SmartLead User Manual

Table of Contents

Advertisement

BioMonitor SmartLead
The Ultium BioMonitor SmartLead has a usable life of seven years.

15.2 Technical Specifications

Input channels: ECG, Respiration via single electrode pair
Sample Rate: 500/1000Hz
Input Range: +/- 24,000 uV
Analog Output scale factor: 0.208mV/uV
Physical Specifications
SmartLead Dimensions:
o
1.9 cm Width x 1.1 cm Thickness, Length varies slightly

15.3 Environmental Conditions for Storage and Transport

Ambient Temperature: 0C to +38C
Relative Humidity: 10% to 100%
Atmospheric Pressure: 70kPa to 107kPa

15.4 Technical Information

Part 852 BioMonitor (Ultium BioMonitor SmartLead)

Each Ultium BioMonitor SmartLead (part # 852) works in-line with an Ultium EMG sensor (part
#810) Each SmartLead is identified by a unique serial number.
The Ultium BioMonitor SmartLead has 3 patient contact points (applied parts). Two points are
standard pinch receptacles for attachment to disposable electrodes. The third patient contact
point is the Ultium sensor itself. Refer to Appendix A for electrode and Ultium sensor placement.

16 Appendices

Appendix A – BioMonitor Electrode Placement
The BioMonitor SmartLead is designed to simultaneously capture an ECG signal, heart rate,
respiration rate, and R-R Interval. To capture this data, two single electrodes are required and are
to be placed as demonstrated below for optimal results. This placement strategy utilizes the
LEFT lead (labeled - L), RIGHT lead (labeled R) and the Ultium Sensor (labeled –S) which acts
as the ground.
P852 Rev B (April 2019)