Non-Contact Voltage; Frequency (Hz) / Duty Cycle - UEi DL479 Instruction Manual

Ac 600a trms hvac/r clamp meter
Hide thumbs Also See for DL479:

Advertisement

WARNING
• Do not attempt to measure more than 2000µA DC
Features:
DC Low Amps – Test Lead input
Range
Resolution
600µA
0.1µA
2000µA
1µA
• Select NCV mode and move the tip of the clamp meter near voltage source.
Both an Audible and Visual alert will indicate voltage.
• Non-Contact Voltage Detection is used to detect power with sensor located
in the tip of the clamp head indicates positive response with both an Audible
and Visual alert.
• Do not use non-contact voltage detector to determine if there is current on
the wire. Detection operation could be affected by socket design, insulation
thickness, type or other factors.
• Voltage indicator light may also light when voltage is present on the meter's
input jack or from an external interference such as motors, flashlights, etc.
On Voltage
Approx. 25V AC
Disconnect test lead probes from voltage source and meter.
• Press Select button to change between Fahrenheit and Celsius
Features:
Range
-31° to 752°F
-35° to 400°C
Sensor must be thermocouple type.
Stated accuracy does not account for thermocouple accuracy.
• Use CAT III test leads or higher.
• Set Function dial to AC/DC Amps, select AC Amps, press and hold the
Select button for Frequency and Duty Cycle modes.
• WARNING
• Do not attempt to measure more than 750V AC/600V DC.
Features:
Frequency
Range
Resolution
99.99Hz
0.01Hz
999.9Hz
0.1Hz
9.999kHz
0.001kHz
Duty Cycle
Range
0.5% to 95%
60Hz to 400Hz
15% to 85%
400Hz to 2kHz
Frequency Width: 60Hz to 400Hz: 0.5% to 95%
Sensitivity: >6Vpp RMS
Accuracy
±(1.2% +3 dgts)

Non-Contact Voltage

True RMS
MAX/MIN
Hz / Duty
Temperature C˚/F˚
True RMS
MAX/MIN
Hz / Duty
Resolution
Accuracy
0.1°F
±(1.5% +3.6°F )
0.1°C
±(1.5% +2 °C)

Frequency (Hz) / Duty Cycle

True RMS
MAX/MIN
Hz / Duty
Accuracy
±(0.1% +3 dgts)
Resolution
0.1%
400Hz to 2kHz: 15% to 85%
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
NCV Sensor
located in tip
• Default =
• Press x1 =
• Default =
• Press and hold =
• Press =
Accuracy
± (0.2% per kHz
+2.0% +2 dgts)
TestEquipmentDepot.com
Overload Protection
2000µA/600V RMS
Overload Protection
600V RMS
Overload Protection
600V RMS
Overload Protection
600V RMS