OPERATION
Please make sure that you have already read and fully understand the WARNING section on page 1.
Making Measurements with the AC Current Probe Model MD301
 Connect the black and red terminals to the Voltage DC range of your DMM or voltage measuring
instrument. Select the appropriate voltage range (2V DC range). Clamp the probe around the conductor
to be tested with the arrow pointed toward the load. If the reading is less than 200mA, select the lower
range until you obtain the best resolution. Read the value display on the DMM and multiply it by the
probe ratio (1000/1). (If reading = 459.0mV, the current flowing through the probe is 0.459V x 1000 =
459A AC)
 For best accuracy, avoid if possible, the proximity of other conductors which may create noise.
Tips for Making Precise Measurements
 When using a current probe with a meter, it is important to select the range that provides the best
resolution. Failure to do this may result in measurement errors.
 Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air
gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for
power measurement.
MAINTENANCE:
Warning
 For maintenance use only original replacement parts.
 To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so.
 To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents into
the probe
Cleaning
To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at all times.
Failure to do so may result in error in readings. To clean the probe jaws, use very fine sand paper (fine
600) to avoid scratching the jaw, then gently clean with a soft, oil cloth.
REPAIR AND CALIBRATION
You must contact our Service Center for a Customer Service Authorization number (CSA#). This will
ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA#
on the outside of the shipping container.
(Or contact your authorized distributor)
NOTE: All customers must obtain a CSA# before returning any instrument.
LIMITED WARRANTY
The current probe is warranted to the owner for a period of one year from the date of original purchase
against defects in manufacture. This limited warranty is given by AEMC
distributor from whom it was purchased. This warranty is void if the unit has been tampered with, abused
or if the defect is related to service not performed by AEMC
Full warranty coverage and product registration is available on our website at:
www.aemc.com/warranty.html.
Please print the online Warranty Coverage Information for your records.
®
®