DESCRIPTION
The MD304 (Catalog #1201.37) is designed for use in industrial environments. The hook-shaped jaws
enable the user to "pry" into or "hook" onto cables (will accept 2 x 500 MCM) or even smaller bus bars.
The current output up to 500Arms continuous makes it the perfect tool for measurement with DMMs,
recorders or instruments with current ranges. The Model MD304 offers a 5 ft lead with safety 4mm
banana plugs.
WARNING
These safety warnings are provided to ensure the safety of personnel and proper operation of the
instrument.
• Read the instruction manual completely and follow all the safety information before attempting
to use or service this instrument.
• Use caution on any circuit: Potentially high voltages and currents may be present and may
pose a shock hazard.
• Read the Safety Specifications section prior to using the current probe. Never exceed the
maximum voltage ratings given.
• Safety is the responsibility of the operator.
• ALWAYS connect the current probe to the display device before clamping the probe onto the
sample being tested.
• ALWAYS inspect the instrument, probe, probe cable, and output terminals prior to use.
Replace any defective parts immediately.
• NEVER use the current probe on electrical conductors rated above 600V in overvoltage
category III (CAT III). Use extreme caution when clamping around bare conductors or bus bars.
INTERNATIONAL ELECTRICAL SYMBOLS
This symbol signifies that the current probe is protected by double or reinforced insulation. Use
only factory specified replacement parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user manual before
using the instrument.
This is a type A current sensor. This symbol signifies that application around and removal
from HAZARDOUS LIVE conductors is permitted.
RECEIVING YOUR SHIPMENT
Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify
your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately
with the carrier and notify your distributor at once, giving a detailed description of any damage.
PACKAGING
The AC Current Probe MD304 is shipped with this instruction manual and a product warranty and
registration card. You can register online at www.aemc.com.
INSTRUMENT COMPATIBILITY
The Model MD304 is compatible with any AC ammeter, multimeter, or other current measurement
instruments with an input impedance lower than 0.1. To achieve the stated accuracy, use the MD304
with an Ammeter having an accuracy of 0.75% or better.

Advertisement

loading

Summary of Contents for AEMC MD304

  • Page 1 “pry” into or “hook” onto cables (will accept 2 x 500 MCM) or even smaller bus bars. The current output up to 500Arms continuous makes it the perfect tool for measurement with DMMs, recorders or instruments with current ranges. The Model MD304 offers a 5 ft lead with safety 4mm banana plugs.
  • Page 2 Ø =1.18" (30 mm) Cable max. 2.5 x 0.19" (63 x 5 mm) Bus Bar max. Ø =0.98" (25 mm) Dual Cable max. 7.68" (195 mm) Black 1.34" (34 mm) 2.6" (66 mm) ELECTRICAL SPECIFICATIONS Accuracy*: Primary 250A; Current Range: current 100A 500A...
  • Page 3 Maximum Conductor Size: 1.18” (30mm) Maximum Bus Bar Size: 2.48 x 0.20” (63 x 5mm) ORDERING INFORMATION Envelope Protection: Current Probe MD304 ....Cat.# 1201.37 IP 20 (IEC 529) Accessories: Drop Test: Banana plug adapter 1.5 m (IEC 68-2-32) (to nonrecessed plug) ....Cat.# 1017.45...
  • Page 4 This warranty is void if the unit has been tampered ® with, abused or if the defect is related to service not performed by AEMC Instruments. Full warranty coverage and product registration is available on our website at: www.aemc.com/warranty.html.