generated. Turn the SPAN control so that the
voltmeter indicates 1.000 V.
3. Repeat steps 1 and 2 until the correct output
voltage is obtained.

5-5 Fixed Output Voltage

The output voltage is fixed under the following conditions:
1. During operations other than weighing
(e.g., the display-off state, calibration)
2. During the zeroing operation, with an 3:
During the zeroing operation, with an 0, an 1, an 2:
3. When "-e" (Weighing pan error) is being displayed:
4. When "e" (Overload error) is being displayed:
〜")
an 0, an 1
Model
GX-200
1.000 V
GX-300
1.000 V
GX-400
1.000 V
GX-600
1.000 V
GX-2000
1.000 V
GX-3000
1.000 V
GX-4000
1.000 V
GX-6100
1.000 V
GX-6000
1.000 V
GX-8000
1.000 V
an 2, an 3
Model
1.050 V
GF-200
1.033 V
GF-300
1.025 V
GF-400
1.017 V
GF-600
1.050 V
GF-1200
1.033 V
GF-2000
1.025 V
GF-3000
1.017 V
GF-4000
1.017 V
GF-6100
1.013 V
GF-6000
GF-8000
32
0 V (or 0.2 V)
The previous output value is retained.
0 V (or 0.2 V when the slide switch
set to "0.2V〜")
0 V (or 0.2 V when the slide switch
set to "0.2V〜")
Output voltage is as shown below.
(when the slide switch is set to "1V
an 0, an 1
an 2, an 3
1.000 V
1.000 V
1.000 V
1.000 V
1.000 V
1.000 V
1.000 V
1.000 V
1.000 V
1.000 V
1.000 V
1.050 V
1.033 V
1.025 V
1.017 V
1.009 V
1.050 V
1.033 V
1.025 V
1.017 V
1.017 V
1.013 V