6

Additional Information

6.1

Specifications

If no power supply cord is provided, please select the proper power supply cord for the ELEPO21 by
referring to "Specifications" and "Certified Cord" below:
Warning
Poring Pulse
1 )
High Voltage Mode (automatic change)
Voltage
Pulse Length
Pulse Interval
Number of Pulses
Polarity Switching
Capacitance
2 )
Low Voltage Mode (automatic change)
Voltage
Pulse Length
Pulse Interval
Number of Pulses
Polarity Switching
Capacitance
Transfer Pulse
Voltage
Pulse Length
Pulse Interval
In case you use a non-approved power supply cord for Nepa Gene products,
Nepa Gene can no longer warrant the electrical safety of the device.
801 - 3,000V (increment: 1V)
1,501 - 3,000V
1,001 - 1,500V
150.1 - 200.0V
801 - 1,000V
50 - 999.9ms (increment: 0.1ms)
0~9
On: With polarity switching, OFF: Without polarity switching
55µF
100 - 800V (increment: 1V)
751 - 800V
501 - 750V
301 - 500V
100 - 300V
50 - 999.9ms (increment: 0.1ms)
0~9
On: With polarity switching, OFF: Without polarity switching
995µF
1 - 300V (increment: 1V)
101 - 300V
0.1 - 50.0ms (increment: 0.1ms)
1 - 100V
0.1 - 99.9ms (increment: 0.1ms)
50.0 - 999.9ms (increment: 0.1ms)
0.1 - 5.0ms (increment: 0.1ms)
0.1 - 10.0ms (increment: 0.1ms)
0.1 - 50.0ms (increment: 0.1ms)
0.1 - 99.9ms (increment: 0.1ms)
0.1 - 15.0ms (increment: 0.1ms)
0.1 - 20.0ms (increment: 0.1ms)
0.1 - 30.0ms (increment: 0.1ms)
0.1 - 50.0ms (increment: 0.1ms)