5 Operating Instructions
5.2.4
Parameters of Transfer Pulse mode
Voltage
The voltage of the first Transfer Pulse (increments: 1V)
Pulse Length
The time duration of one Transfer Pulse in milliseconds (increments: 0.1ms)
Pulse Interval
The time interval between Transfer Pulses in milliseconds (increments: 0.1ms)
Number of Pulses
The number of Transfer Pulses
Polarity Switching
This function reverses the polarity of Transfer Pulses. When the "+" indicator of the Polarity selection
option in the Set Parameters frame is illuminated, it means that the polarity switching function is OFF.
When the "+/-"indicator of "the Polarity selection option is illuminated, it means that the polarity switching
function is ON. If the voltage, number of pulses and polarity switching (+/-) are programmed, the out put
Pulses will be as follows;
5.2.5
Setting parameters of Transfer Pulse mode
Setting Voltage, Pulse Length, Pulse Interval, Number of Pulses and Decay
Rate
The same as Poring Pulse mode. See "5.2.3. Setting parameters of Poring Pulse mode" on this manual.
Setting Polarity Switching
1
Select the Polarity Switching option in the Transfer Pulse frame in the Selection window by pressing ◄ ▴
▾ ▸ buttons.
5-4
Time