FEATURES AND FUNCTIONS
•
3 sizes available covering the calibration ranges of:
1.2 to 60 N∙m.
8 to 400 N∙m.
30 to 1500 N∙m.
•
5 digit resolution.
•
ISO 6789-1:2017 limit mode with single press selection for 20%, 60% & 100% target values.
•
NON ISO 6789-1:2017 limit mode with user defined tolerances.
•
The pictorial keys allow direct access to measurement modes.
•
Serial Port Connector for data output to PC's or printers.
•
PRINT/NO PRINT ( PRINT/X PRINT) feature for control of serial port output.
•
The last 5 readings taken are memorised on the display.
•
SET UP menu's for:
1.
Language of operation,
2.
Limits mode required,
3.
Units of measurement enabled/disabled,
4.
Serial port communication parameters,
Thresholds for 'CLICK & CAM' mode,
5.
6.
Modes of measurement enabled/disabled.
4