Keysight Technologies N5252A User Manual page 14

Vector network analyzer system
Hide thumbs Also See for N5252A:
Table of Contents

Advertisement

Port 2 Configuration
1. Install the two cables between M9374A Module #2 and VDI Extender #2. See
Figure
6.
2. If not already installed, install two 50 Ω loads to LO In and LO Out on M9374A Module #2. If
not already installed, install a reference jumper and a trigger jumper on Modules #1 and #2.
See
Figure
6.
3. Add a 50 Ω load (use the load removed from Module #1) to LO-O on VDI Extender #2. See
Table 4
and
Figure
Extender #3 (LO-I) with an RF/LO cable.
4. Connect the power cord to VDI Extender #2. See
An RF cable up to 5 meters in length can be used for an LO-O to LO-I connection
between extenders. The M9374A LO Out can only support 1.2 meter cable lengths.
Table 4
Port 2 Configuration Cable Connections
Figure
Cable
Type
Reference
A.
RF/LO
B.
IF
50 Ω Load (or RF/LO
C.
for Option 400)
Figure 6
Port 2 Configuration
14
6. If Option 400 is to be used, connect VDI Extender #2 (LO-O) to VDI
M9374A
Module #2
Port 1
Port 2
-
A.
Figure
6.
to
to
M9374A Module #2 with VDI Extender #2
C.
Load (or RF/LO cable to LO-I of VDI Extender #3)
B.
Table 4
and
VDI
Extender #2
RF/R
M
LO-O
User's Guide N5252-90002