1
Getting Started

Electrical Connectors

C A U T I O N
32
Figure 23
Polarization maintaining fiber
There are two BNC connectors on the rear panel of your instrument.
These are the Trigger Out and the Trigger In connectors.
The N7747A and N7748A have a front panel BNC connector for each
channel. These are for analog output signals and provide a voltage
proportional to the input signal power in mW. The voltage is in the range
0-2 V and 1 V corresponds to the value of the currently selected power
range. For example, when using the -10 dBm range, a 100 µW optical
power will give 1 V analog output.
The Trigger In is a TTL input. A maximum of 5 V can be applied as an external voltage
to this input connector.
The Trigger Output is a TTL output. Do not apply an external voltage to this
connector.
No external voltage should be applied to the analog output connectors of the
N7747A and N7748A.