Recommended separation distances between portable and mobile
RF communications equipment and the EQUIPMENT or SYSTEM –
for EQUIPMENT or SYSTEM that are not LIFE-SUPPORTING
portable and mobile RF communications equipment and the ECG100G ECG
The ECG100G ECG is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the ECG100G ECG can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the ECG100G ECG as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
User Manual
Recommended separation distances between
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
3
5 .
d
P
V
1
0.117
0.369
1.17
3.69
11.7
27
(m)
80 MHz to 800 MHz
3
5 .
d
P
E
1
0.117
0.369
1.17
3.69
11.7
800 MHz to 2.5 GHz
7
d
P
E
1
0.233
0.738
2.333
7.379
23.33